111
2920B EN 20031206
Establishing the daily operating reference density levels
This procedure enables you to establish the base line values for:
•
Base+Fog density,
•
Mid density,
•
DD (Density Difference),
•
MaxD (maximum density).
To establish the daily operating density levels, proceed as follows:
1
Press the Key-operator key to enter the key-operator mode.
2
Press the Down key seven times, followed by the Confirm key to select
‘Quality control’.
A confirmation screen appears:
3
Press the Confirm key (YES) to start the QC procedure or the Escape key
(NO) to quit.
The Drystar 5300 will automatically print the QC test image.
By default, the internal densitometer of the Drystar 5300 will be used. The
internal densitometer of the Drystar 5300 is calibrated at installation.
Authorized service personnel should recalibrate the internal
densitometer annually or after major service or repair.
Do not change the ‘default’ internal densitometer during the entire QC
procedure!
1 Show settings
2 Change settings
3 Print image
4 Save configuration
5 Restore config.
6 Calibration
7 Service actions
8 Quality control
9 Installation
KO
Start QC
Procedure
YES (v)
NO (x)
QC
QC procedure
in progress
Please wait
62%
QC
Содержание Drystar 5300
Страница 1: ...Drystar 5300 Reference manual POWER ...
Страница 154: ...154 2920B EN 20031206 4 Lower the film positioner unit 5 Remove the film positioner cable out of its retainers ...
Страница 156: ...156 2920B EN 20031206 8 Using the cross head screwdriver remount the 3 screws to fix the film position unit ...
Страница 160: ...160 2920B EN 20031206 ...
Страница 168: ...168 2920B EN 20031206 ...
Страница 189: ...Equipment information sheet Appendix A ...
Страница 195: ...Drystar media density response data Appendix B ...
Страница 198: ...198 2920B EN 20031206 ...
Страница 199: ...Quality Control Charts Appendix C ...
Страница 205: ...Index Appendix D ...
Страница 210: ...Printed in Belgium Published by Agfa Gevaert N V B 2640 Mortsel Belgium 2920B EN 20031206 ...