FAILURE MESSAGES (TO BE CONTINUED)
CARRIAGE EOT +
This indicates a system failure with the carriage end position on foot side and
forbids all the movements
Press OK.
- If the message does not
appear again, move the
system slightly in the
other way and try again
the movement.
Then, if the message
appears, turn the system
OFF and call Field
Service.
- If the message appears
again, turn the system
OFF and call Field
Service
CARRIAGE EOT -
This indicates a system failure with the carriage end position on head side and
forbids all the movements
COLUMN EOT +
This indicates a system failure with the column incidence end position on foot
side (+40°)and forbids all the movements
COLUMN EOT -
This indicates a system failure with the column incidence end position on head
side (-40°)and forbids all the movements
TILTING EOT +
This indicates a system failure with the table tilting end position on foot side
(+90°)and forbids all the movements
TILTING EOT -
This indicates a system failure with the table tilting end position on head side
(-25°) and forbids all the movements
TABLE HEIGHT EOT +
This indicates a system failure with the table variable height in higher position
and forbids all the movements
TABLE HEIGHT EOT -
This indicates a system failure with the table variable height in lower position
and forbids all the movements
LONG TABLE TOP EOT +
This indicates a system failure with the tabletop longitudinal end position on
foot side and forbids all the movements
LONG TABLE TOP EOT -
This indicates a system failure with the tabletop longitudinal end position on
head side and forbids all the movements
S.I.D. EOT +
This indicates a system failure with the tube movement (SID) in higher position
and forbids all the movements
S.I.D. EOT -
This indicates a system failure with the tube movement (SID) in lower position
and forbids all the movements
LAT TABLE TOP EOT +
This indicates a system failure with the tabletop lateral end position on front
side and forbids all the movements
LAT TABLE TOP EOT -
This indicates a system failure with the tabletop lateral end position on rear side
and forbids all the movements
BACK TO AUTOPOSITIONING
This indicates a system failure with the autopositioning and forbids all the
movements
MANUAL COLLIMATOR
This indicates a system failure with the automatic collimator and forbids all the
movements
Stitching not Possible, Move Carriage
Regarding the number of exposures selected for stitching, the carriage position is
too close to end of table.
Move the carriage up
and try again the
acquisition
Tomography not Possible, Move Carriage
Regarding incidence and number of exposures selected for tomography, the
carriage position is too close to end of table.