Digital Remote Controlled Table with dynamic flat panel detector
User’s Manual
35
INFORMATION MESSAGES (TO BE CONTINUED)
Messages displayed on top of screen
description
What to do
TABLE TOP REAR LIMIT
This indicates that the tabletop reaches its end lateral position on rear side (column)
These messages are
informative and do not
require pressing any key.
These indications stay
displayed until any
change (other control ...).
If the message persists
when you order another
movement, turn the
system OFF and ON.
If the equipment remains
inoperative, turn it OFF
and call Field Service
TABLE TOP FRONT LIMIT
This indicates that the tabletop reaches its end lateral position on front side
ELEVATION LOWER LIMIT
This indicates that the tabletop reaches its lower position
ELEVATION UPPER LIMIT
This indicates that the tabletop reaches its higher position
Collimator Not Ready
This indicates that the collimator is opening or closing to the desired format
Compressor Upper limit
This indicates that the compression device reaches its higher position
Compressor Lower limit
This indicates that the compression device reaches its lower position
Tomography In Progress
This indicates that the tomography device is moving
Patient Compression Detected
This indicates that the system detects compression
Max. Compression Level Reached
This indicates that the maximum compression level is reached
Patient Compression Detected, Tube
Angulation Disabled
This indicates that the system detects compression and then forbids any
movement of angulation
In order to allow the
desired movement, lift UP
the compression device
Patient Compression Detected, Carriage
Disabled
This indicates that the system detects compression and then forbids any
movement of carriage
Patient Compression Detected, Lat. Table
Top Disabled
This indicates that the system detects compression and then forbids any lateral
movement of tabletop
Patient Compression Detected, Long.
Table Top Disabled
This indicates that the system detects compression and then forbids any
longitudinal movement of tabletop
TUBE ROTATED, FLUORO
DISABLED
This indicates that the X-ray tube is rotated and then not centred on the detector
Place X-ray tube faced to
detector, at 0 ° with the
indexing system.
NO AUTO POSITIONING SELECTED
This indicates that you did not select any anatomical programme so then no
autopositioning before pressing the MOVE button
Select an anatomical
programme before
wanting to position the
system by MOVE button
ELEVATION LOWER LIMIT:
TABLE IS TILTED
This indicates that the system cannot go lower the tabletop as table is tilted
To reach elevation lower
limit, untilt the table
before lowering
Column Angulation Is Not Enabled
For SID > 150 cm
This indicates that you are not allowed to make a column incidence if the SID is
over 150 cm
Decrease the SID under
150 cm to be able to make
an incidence
Position Reached
This indicates that the system reaches the selected position
Now, you can follow your
exam procedure
End of Stitching
This indicates that the system has just finished the last exposure of the stitching
programme
Now, you can end your
exam procedure
End of Tomography
This indicates that the system has just finished the last exposure of the
tomography programme
Now, you can end your
exam procedure
ROTATION IMPOSSIBLE,
PRESS MOVE
(for version with removable detector)
This indicates that the system cannot rotate the detector due to its support
position and selected programme
For version with
removable detector.
Press MOVE to cancel
carriage