The X-ray system cannot be moved
Details
The X-ray system cannot be moved.
First
problem
situation
Fault in the exposure button (wired or wireless).
An error is displayed on the software console: “Exposure switch
error”.
Possible causes are damage from falling or liquid having entered
the exposure button, e.g. after cleaning or disinfecting.
Solution • Wired exposure button: detach the exposure button and re-
start DR 100s.
• Wireless exposure button: remove the battery from the expo-
sure button and restart DR 100s.
Second
problem
situation
Failure of battery or motor.
Solution
CAUTION:
Never apply force on the handlebar to move the
equipment. Two persons are required to move the
X-ray system.
1.
Put the equipment in secured position.
2.
Turn OFF the equipment.
3.
Identify the red metallic release cable (A), placed under the
control panel of the x-ray generator. For the movement of the
equipment, pull and keep pulled downwards the metallic ca-
ble.
Figure 65: Red cable to release movement
4.
Move the equipment with two people. One person in the front
to pull and drive the equipment, the other one in the back to
keep the red metallic unlocking cable and push the equip-
ment.
148
| DR 100s | Problem solving
0411D EN 20220627 1040