4.
Turn on the detector.
Note
:
Before operating the detector, start up the NX
workstation.
Press and hold the power button for 1 second.
Figure 14: Power button
After startup the green status indicator is lit, indicating the power status.
5.
Turn on the System Control Unit using the power switch.
The status indicator is green.
Depending on the configuration, the System Control Unit may not be part
of the system.
6.
Check the DR Detector Status icon on the DR Detector Switch.
If the displayed status is error, register the DR Detector to the NX
workstation.
If the DR Detector communicates via the internal wireless adapter of the
workstation, switch the NX Workstation to the wireless network of the DR
Detector.
The green and orange status indicators on the DR Detector are lit. The DR
Detector is ready.
Before exposure make sure to check the equipment daily and confirm that it
works properly.
Related Links
on page 90
Registering the DR Detector on another NX Workstation
Using the Windows wifi settings to switch between the wireless DR detector and
on page 94
DR detector not ready for exposure