Registering the DR Detector on another NX
Workstation
The DR Detector can be used for examinations on different NX workstations.
The DR Detector is set up to communicate to a specific NX workstation. The
procedure of registering the DR Detector to another NX workstation switches
the availability of the DR Detector between NX workstations.
In a configuration with multiple NX workstations sharing a DR Detector, each
NX workstation is connected to a System Control Unit. One or more of the
System Control Units is equipped with a DR Detector Cable.
In a configuration with mobile X-ray units sharing a DR Detector, a dedicated
PC is connected to a System Control Unit that is equipped with a DR Detector
Cable.
Note
:
On a PC without NX software, the VDI Panel Sharing Tool
may not be set up to start automatically. To start it manually, in
the Start menu, select All programs > Agfa > Start VDI Panel
Sharing Tool.
To register the DR Detector and make a connection to another X-ray room:
1.
Connect the DR Detector to any of the NX workstations with the DR
Detector Cable.
A dialog is displayed on the NX workstation listing the configured X-ray
rooms.
It may take up to 30 seconds for the dialog to pop up.
2.
Select the X-ray room where the DR Detector will be used.
A dialog is displayed on the NX workstation to confirm the registration.