Performing quality control (QC) tests for mammography
application
The following procedures must be performed daily, weekly or annually as
indicated. When used in a mobile application it is required to perform QC
after every move.
The reason for performing quality control tests is to determine if any
significant image quality variation or deterioration has occurred which may
require corrective action. Comparing the results of the tests with the reference
values previously established does this.
This procedure allows the operator to take the necessary preventive actions
before any image quality loss can take place.
Topics:
•
Performing the daily QC test for mammography application
•
Performing the Weekly QC tests for mammography application
•
Performing the Annual QC tests for mammography application
Performing the daily QC test for mammography application
WARNING:
This test must be performed every day before any clinical film
can be processed.
1.
Turn on the printer and wait at least for 15 minutes.
2.
Press the Key-operator key to enter the Key-operator mode.
The ‘Key-operator main menu’ screen appears:
3.
Press the Down key eight times, followed by the Confirm key to select
‘Quality Control’.
The ‘Select input tray’ screen appears:
106
| DRY 3.4 | Advanced Operation (key-operator mode)
2602A EN 20191018 0826
Содержание 5364/500
Страница 1: ...DRY 3 4 5364 500 User Manual 2602A EN 20191018 0826 ...
Страница 12: ...12 DRY 3 4 DRY 3 4 2602A EN 20191018 0826 ...
Страница 35: ...Compliance Topics General Safety Electromagnetic Compatibility DRY 3 4 Introduction 35 2602A EN 20191018 0826 ...
Страница 58: ... Audio signals The keypad The display 58 DRY 3 4 Introduction 2602A EN 20191018 0826 ...
Страница 87: ...The menu structure DRY 3 4 Advanced Operation key operator mode 87 2602A EN 20191018 0826 ...
Страница 125: ... Cables transducers and accessories DRY 3 4 Remarks for HF emission and immunity 125 2602A EN 20191018 0826 ...
Страница 160: ...4 Open the upper lower input tray 160 DRY 3 4 Plug Play Installation Manual 2602A EN 20191018 0826 ...