accordance with
IEC 61000-4-8
the typical val-
ues as they are
in a
commercial and
clinical environ-
ment.
REMARK : U
r
is the alternating current in the network before the application
of the test level.
This device is intended for operation in the electromagnetic environment
given below. The user of the device should ensure that it is used in such
an environment.
Tests of Resist-
ance to Dis-
ruption
IEC 60601-1-2:2014
Test Level
Level of
Agree-
ment
Electromagnetic En-
vironment
Use portable and mo-
bile radio sets at a
safe distance from the
device (including the
leads) not closer than
the recommended
protective distance,
which is calculated
according to the equa-
tion suitable for the
RF emission frequen-
cy.
Recommended pro-
tective distance:
Conducted
high frequency
disturbance
variables in ac-
cordance with
IEC 61000-4-6
6 V/m in ISM fre-
quency bands be-
tween 150 kHz and
80 MHz
6 V/m in
ISM fre-
quency
bands be-
tween 150
kHz and 80
MHz
d
= 1.2
Radiated high
frequency dis-
turbance varia-
bles in accord-
ance with IEC
61000-4-3
10 V/m
80 MHz to 2.7 GHz
10 V/m
80 MHz to
2.7 GHz
d
= 1.2
80 MHz
to 800 MHz
122
| DRY 3.4 | Remarks for HF-emission and immunity
2602A EN 20191018 0826
Содержание 5364/500
Страница 1: ...DRY 3 4 5364 500 User Manual 2602A EN 20191018 0826 ...
Страница 12: ...12 DRY 3 4 DRY 3 4 2602A EN 20191018 0826 ...
Страница 35: ...Compliance Topics General Safety Electromagnetic Compatibility DRY 3 4 Introduction 35 2602A EN 20191018 0826 ...
Страница 58: ... Audio signals The keypad The display 58 DRY 3 4 Introduction 2602A EN 20191018 0826 ...
Страница 87: ...The menu structure DRY 3 4 Advanced Operation key operator mode 87 2602A EN 20191018 0826 ...
Страница 125: ... Cables transducers and accessories DRY 3 4 Remarks for HF emission and immunity 125 2602A EN 20191018 0826 ...
Страница 160: ...4 Open the upper lower input tray 160 DRY 3 4 Plug Play Installation Manual 2602A EN 20191018 0826 ...