To identify artifacts and verify spatial resolution you must perform the
following test weekly or as needed for troubleshooting image quality
problems.
1.
First, print out the QC mammography test image.
2.
Check the test image visually for artifacts: no significant disturbing
artifacts should be visible.
3.
Check the spatial resolution.
The test film also shows three squares which each contains an oval. These
3 ovals contain 3 groups, each having 5 dots. All five dots of each group
must be visible with a magnifying glass. The smallest cluster of 5 dots is
only visible if the viewing conditions are good.
4.
Check the Low Contrast Visibility at both the high (100 / 95%) and low
end
(0 / 5%) of the density scale. You should be able to see the circle in the
square (refer to item 1 on the
‘QC mammography test image’
) and the
upper circle (refer to item 2 on the
‘QC mammography test image’
).
5.
Record these values on Chart 3 (‘Artifacts and Spatial Resolution Control
Chart’).
WARNING:
In case of significant artifacts, insufficient spatial resolution
or failure of any other recommended QC tests, the cause of
the problem must be identified, and corrective action must
be taken before the printer can be used for any further
clinical imaging. If the problem can not be solved, call your
local service organization for assistance.
Related Links
Performing the daily QC test for mammography application
Establishing the image geometry reference values for mammography application
on page 103
on page 135
Preventive maintenance schedule
on page 111
Performing the Annual QC tests for mammography application
Geometric Consistency Test
To be able to notice fluctuations in image size and aspect ratio, you must
perform this procedure once a year.
1.
First, perform the daily test.
2.
Measure the distances A and B of the geometric square on the QC
mammography test image.
108
| DRY 3.4 | Advanced Operation (key-operator mode)
2602A EN 20191018 0826
Содержание 5364/500
Страница 1: ...DRY 3 4 5364 500 User Manual 2602A EN 20191018 0826 ...
Страница 12: ...12 DRY 3 4 DRY 3 4 2602A EN 20191018 0826 ...
Страница 35: ...Compliance Topics General Safety Electromagnetic Compatibility DRY 3 4 Introduction 35 2602A EN 20191018 0826 ...
Страница 58: ... Audio signals The keypad The display 58 DRY 3 4 Introduction 2602A EN 20191018 0826 ...
Страница 87: ...The menu structure DRY 3 4 Advanced Operation key operator mode 87 2602A EN 20191018 0826 ...
Страница 125: ... Cables transducers and accessories DRY 3 4 Remarks for HF emission and immunity 125 2602A EN 20191018 0826 ...
Страница 160: ...4 Open the upper lower input tray 160 DRY 3 4 Plug Play Installation Manual 2602A EN 20191018 0826 ...