4.2 Maintenance overview
Table 4.2 Maintenance Procedures
4.3 Pin soak
It is recommended that the pins be left submerged in the wash solution whenever the
instrument is not in use. This keeps the pins wet and aids in dispensing accuracy. The pin
soak must be manually performed by the operator.
To soak the pins:
1.
Go to the Maintenance screen (tap the Back button
on any screen and tap the
MAINTENANCE
button
on the Main Menu).
2.
Tap
CLEAN DAILY
on the Maintenance screen
The pins are immersed in the wash solution.
3.
In the daily pin cleaning dialog that appears, tap
SOAK
.
The timer stops counting down, allowing the pins to remain in the wash station.
Tapping the
SOAK
button again will cause the timer to continue the countdown;
tapping the
CANCEL
button aborts the cleaning operation.
Perform
Description
See Page
Pin Soak
It is recommended that the pins be submerged in
the wash solution whenever the instrument is not
in use.
Daily
Clean the pins.
Drain and fill the ultrasonic wash supply bottle.
Flush the calibrant reservoir.
Fill the supply tank.
Drain the waste tank.
Monthly (only needed if
processing fewer than 10
SpectroCHIP Arrays per month)
Clean the supply and waste tanks and flush the
wash system.
Quarterly (or as directed by
Agena BIoscience Customer
Support)
Condition the pins.
Quarterly (or every 100
SpectroCHIP Arrays processed,
whichever comes first)
Drain the vacuum tank.