Chapter 4 Maintenance Procedures
94
USG-CUS-059 Rev05
MassARRAY® Nanodispenser RS1000 v2.1 User Guide
For Research Use Only. Not for use in diagnostic procedures.
9.
Repeat the Sonicator Solution drain
. Tap
SONICATOR SOLUTION Drain
on the Maintenance screen.
This drains the ethanol from the ultrasonic wash. An on-screen message displays the
time remaining for the operation. Wait for draining to complete.
Drain and fill the
ultrasonic wash supply
bottle
Drain and fill the ultrasonic wash supply bottle at the beginning of each day of use. A full
ultrasonic wash supply bottle holds approximately 50 mL. This should be enough for a
day of instrument use.
NOTE
The instrument should be left in pin soak when not in use (
). However, in the event that
the instrument is not in pin soak, the ultrasonic wash and its supply bottle will be automatically
drained at 1:00 AM (01:00) every morning.
If
Figure 4.6 Starting the Daily Pin Cleaning
Figure 4.7 Ultrasonic Wash Supply Reservoir
Ultrasonic wash
supply bottle
Wash station block