AffloVest RP8001 Скачать руководство пользователя страница 16

E

LECTROMAGNETIC

E

NVIRONMENT

G

UIDELINES

I

NTERFERENCE

I

MMUNITY

T

ESTS

IEC 60601-T

EST

L

EVEL

C

OMPLIANCE

L

EVEL

Portable and mobile RF communications equipment
should be used no closer to any part of the
AffloVest, including the cables, than the
recommended distance calculated from the
equation applicable to the transmit frequency.
Recommended distance:
d = 1.2 m P

d=1.2 m P for 80 MHz to 800 MHz
d=2.3 m P for 800 MHz to 2.5 GHz

Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended distance in
meters (m). The field strength of fixed RF
transmitters, as determined by an electromagnetic
site survey

a

should be less than the compliance

level in each frequency range

b

. Interference may

occur in the vicinity of equipment marked with the
following symbol.

3 V

rms

3 V/m

3 V

rms

150 kHz to
80 MHz

3 V/m
80MHz to
2.5 GHz

Conducted RF
interference as
per IEC 61000-4-6

Radiated RF
interference as
per IEC 61000-4-3

NOTE 1: The higher frequency range applies at 80 MHz and 800 MHz.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorbtion and reflection
from structures, objects, and persons.

a

The field strength from fixed transmitters, such as base stations for radio telephones and mobile terrestrial radio equipment,

amateur radio stations, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess
the electromagnetic environment in terms of fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the AffloVest is used exceeds the applicable compliance level specified above,
the AffloVest should be observed to verify normal operation. If abnormal performance features are observed, additional
measures may be necessary, such as re-orienting or relocating the AffloVest.

b

The field strength should be less than 3 V/m over the frequency range from 150 kHz to 80 MHz.

The AffloVest is intended for use in the electromagnetic environment specified below. The customer or user of the
AffloVest should ensure that it is used in such an environment.

Содержание RP8001

Страница 1: ......

Страница 2: ...it Charger RP8001 Rechargeable Battery RP8003 AffloVest L RP8300 M RP8200 S RP8100 Carrying Case RP8005 AffloVest Instructions for Use Adapter Cable for D C Operation RP8002 g y Aff ff f loVe Ve V st...

Страница 3: ...place for future reference With proper maintenance and care you will be rewarded by your AffloVest with a long service life and trouble free operation of our product We greatly value the satisfaction...

Страница 4: ...Charger Test and Identification Symbols RP8003 Battery Pack Not for Outdoor Use T V Rheinland Approval In Accordance with EN60601 1 and IEC 60601 1 Communan Europ enne European Community Directives g...

Страница 5: ...A Nominal Voltage Nominal Power Consumption Power Supply Unit Charger Lithium Polymer Battery Pack Manufacturer Distributor Sizes 12V DC Approx 30W Model RP 8001 INPUT 100 240V AC 50 60 Hz OUTPUT 12V...

Страница 6: ...proceduresonorinjuriestothespinal column The AffloVest is intended for promoting airway clearance and improvement of bronchial drainage by enhancing mobilization of bronchial secretions where external...

Страница 7: ...ry has reached its maximum charge disconnect it from the charger and connect it to the AffloVest It is not recommended to start a new cycle if more than half of the LED lights on the battery are not i...

Страница 8: ...nductive surface Lithium Polymer batteries that are totally discharged are defective and must no longer be used Do not expose the battery pack to direct sunlight The maximum permissible ambient temper...

Страница 9: ...level of intensity Programs P1 P2 and P3 always start with SOFT intensity Note The intensity automatically switches to SOFT when changing the programs P1 P2 and P3 You can then freely select your chos...

Страница 10: ...functions are available again after pressing any button The LED lights P1 P2 or P3 on the hand held controller light up to indicate the currently active program The LED lights 1 5 show which oscillati...

Страница 11: ...ns The use of any other power supply units and chargers can lead to fire and electric shock Overcharging must be avoided Pay attention to the voltage ratings on the power supply unit The device is des...

Страница 12: ...epairs shall not extend the warranty period Incorrect use or operation e g incorrect charging or excessively high discharge rates shall invalidate any guarantee claims The same applies to defects caus...

Страница 13: ...ailed information at your nearest collection point for recyclable materials from your local authority Recycling instead of waste disposal Devices accessories and packaging should be recycled only for...

Страница 14: ...ly fuse defective or automatic circuit breaker tripped Power Failure Standby Mode Plug in correctly Plug in power plug and or Plug in charge battery Needs to be repaired by an IBC authorized electrici...

Страница 15: ...ive humidity should be at least 30 Fast Electrical Transients Bursts as per IEC 61000 4 4 2 kV for Power Supply Lines 1 kV for Input Output Lines Mains power quality should be that of a typical commer...

Страница 16: ...The higher frequency range applies at 80 MHz and 800 MHz NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorbtion and reflection from structures o...

Страница 17: ...Transmitter Output W 150 kHz to 80 MHz d 1 2 m P 80 MHz to 800 MHz d 1 2 m P 800 MHz to 2 5 GHz d 2 3 m P 0 01 0 1 1 10 100 0 12 1 2 0 38 3 8 12 0 12 0 23 0 38 0 73 1 2 2 3 3 8 7 3 12 23 The recommend...

Отзывы: