background image

2

micro-Line saggital saw GD307

Aesculap Power Systems

Copyright information

Copyright© 2006
Aesculap AG & Co. KG®
All rights reserved 11/06

This service information is copyrighted. This service information must not
be copied or reproduced by any means either in whole or in part without
the express permission of Aesculap AG & CO. KG.

This manual is intended for informational purposes only. Ownership of this
manual alone does not constitute or imply authorization to service the
product.

The technical information, illustrations and dimensions contained in this
manual are non-binding. No claims may be made on the basis of the
information contained herein. We reserve the right to make improvements
without altering this documentation. We reserve the right to make
technical changes without prior notice.

Main address for servicing:

Aesculap Technischer Service (ATS)
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone: +49 

7461 

95-2700

Fax:

+49 7461 16-2887

E-mail:

[email protected]

Other service addresses can be obtained through the address indicated
above.

Manufacturer’s liability

We expressly point out that we can accept responsibility for any effects
on the safety, reliability, and performance of our medical products, if and
only if:

• any assembly, extensions, readjustments, modifications, or repairs are

performed by technically experienced, knowledgeable and trained
personnel and

• the medical products are used as set forth in the instructions for use.

It is possible to learn servicing the respective medical products through
appropriate courses of instruction offered by Aesculap. To arrange for such
a course, please contact Aesculap Technical Service (ATS).

To ensure that your Aesculap warranty remains valid, we recommend
using for repairs only Aesculap factory-tested spare parts. Spare parts, as
well as the relevant tools, can be ordered from Aesculap Technical Service
(ATS). Any unauthorized opening and/or alterations of the medical product
by third parties lead to the exclusion of our liability, as far as a fault is
attributable to such unauthorized opening and/or alteration of the
product. Aesculap cannot accept responsibility for the use of unsuitable
spare parts, tools or devices. After repairs, dropping, severe damage or
misuse, the product should be inspected by a qualified person.

This manual refers to the product as it was when the manual was
prepared. Technical changes, particularly of software, may be made at any
time.

General advisory

This service manual contains illustrations and explanations. Each
explanation covers the following: modes of operation, parts list, assembly,
disassembly, functionality test, and a tools/devices section (if applicable).

For further information about a specific product (e.g., troubleshooting list,
accessories) please refer to the relevant user instructions.

It is crucial to maintain extreme cleanliness during repairs of the products
described below. All products must undergo thorough function tests after
any repairs. If, in the course of a repair, a labeled component is replaced,
the labeling must be transferred to the spare part. Either the original label
should be used or the spare part bearing the relevant label should be
ordered. Any such order should include the relevant information.

Unless otherwise indicated, all medical products should be inspected
annually.

General information about motors and handpieces

It is critically important that ball bearings be inspected annually. Even the
slightest defect or dirt can lead to overheating during use, which would
render the product unusable. If for the repair of e.g. a motor a ball bearing
has to be replaced, it is best in most cases to also replace all other ball
bearings. 

If it is necessary to heat up adhesive joints during disassembly, a hot-air
heat gun is the appropriate tool. The smallest possible nozzle should be
used so as to avoid damaging other components. For glue joints, make
certain that the surfaces to be joined are absolutely clean and free of
grease. We recommend using Loctite Rapid Cleaner 7063 for this purpose
(WS. no. 520001750). Only use adhesives listed in this manual. Follow the
glue manufacturer’s recommendations.

Lubricate components with the recommended Aesculap product only.

Following disassembly, clean all components thoroughly and inspect them
for any damage. Do not install any component of uncertain condition.
Products must always be test-run for several minutes after any repair.

Documentation

Art. no.

Instructions for use

TA009539

Содержание GD307

Страница 1: ...Aesculap Power Systems Service Manual micro Line saggital saw GD307 Service Manual micro Line Sagittals ge GD307...

Страница 2: ...TA003323 GD307200 GD307218 TA009542 TA009541 GD307219 GD307221 GD307222 GD307223 GD307202 GD307204 GD307224 GD307225 GD307226 GD307227 GD307228 GD307229...

Страница 3: ...GD307207 GD307205 GD307210 TA007307 GD307803 GD307201 GD307802 GD307216 GD307217 TA007306 GD307203 GD307228 GD307227...

Страница 4: ...y for the use of unsuitable spare parts tools or devices After repairs dropping severe damage or misuse the product should be inspected by a qualified person This manual refers to the product as it wa...

Страница 5: ...inspection 7 9 Parts list 7 1 Safe handling Operate the micro Line saggital saw GD307 only with authentic Aesculap accessories 2 Tools auxiliary materials supplies 2 1 Tools 2 2 Auxiliary materials 2...

Страница 6: ...autiously saw into a block of wood Check the micro Line saggital saw for atypical running noises that might be a sign of defective bearings gear wheels or connections Check the micro Line saggital saw...

Страница 7: ...ving Steel ball jammed Clean or lubricate steel ball 2 mm NR TA003323 Motor running but not driving the saw Gearing defective Replace drive unit complete GD307802 Insufficient cutting power of saw bla...

Страница 8: ...s completes disassembling of the rear housing 6 3 Disassembling the drive unit complete Expel pin GD307216 from drive unit complete GD307802 Take off coupling piece GD307228 and compression spring GD3...

Страница 9: ...Cautiously saw into a block of wood Check the micro Line saggital saw for atypical running noises which would be a sign of defective bearings gear wheels or connections Check the micro Line saggital...

Страница 10: ...starkem Schaden oder Missbrauch des Produkts Dieses Manual entspricht dem Stand bei seiner Erstellung Mit techni schen nderungen muss jederzeit gerechnet werden insbesondere bei Software Grundlegende...

Страница 11: ...rfl chen pr fen 13 9 St ckliste 13 1 Sichere Handhabung micro Line Sagittals ge GD307 nur mit Original Aesculap Zubeh r betreiben 2 Werkzeuge Hilfsmittel Betriebsstoffe 2 1 Werkzeuge 2 2 Hilfsmittel 2...

Страница 12: ...cro Line Sagittals ge vorsichtig in ein Holzst ck s gen micro Line Sagittals ge auf untypische Laufger usche pr fen die auf defekte Lager Zahnr der und Verbindungen schlie en lassen micro Line Sagitta...

Страница 13: ...bel bewegt sich nicht Stahlkugel sitzt fest Stahlkugel 2 mm NR TA003323 reinigen bzw len Antrieb dreht sich allein Defektes Getriebe Antrieb komplett GD307802 ersetzen Ungen gende Schnittleistung der...

Страница 14: ...sehinterteil pressen Das Geh usehinterteil ist demontiert 6 3 Antrieb komplett demontieren Stift GD307216 aus dem Antrieb komplett GD307802 schlagen Kupplungsteil GD307228 und Druckfeder GD307217 abzi...

Страница 15: ...ittals ge vorsichtig in ein Holzst ck s gen micro Line Sagittals ge auf untypische Laufger usche pr fen die auf defekte Lager Zahnr der und Verbindungen schlie en lassen micro Line Sagittals ge auf Er...

Страница 16: ...tlingen Germany Phone 49 7461 95 0 Fax 49 7461 95 2600 www aesculap de TA Nr 011235 10 06 nd Nr CE Kennzeichnung gem Richtlinie 93 42 EWG CE marking according to directive 93 42 EEC Technische nderung...

Отзывы: