micro-Line saggital saw GD307
Aesculap Power Systems
Copyright information
Copyright© 2006
Aesculap AG & Co. KG®
All rights reserved 11/06
This service information is copyrighted. This service information must not
be copied or reproduced by any means either in whole or in part without
the express permission of Aesculap AG & CO. KG.
This manual is intended for informational purposes only. Ownership of this
manual alone does not constitute or imply authorization to service the
product.
The technical information, illustrations and dimensions contained in this
manual are non-binding. No claims may be made on the basis of the
information contained herein. We reserve the right to make improvements
without altering this documentation. We reserve the right to make
technical changes without prior notice.
Main address for servicing:
Aesculap Technischer Service (ATS)
Am Aesculap-Platz
78532 Tuttlingen / Germany
Phone: +49
7461
95-2700
Fax:
+49 7461 16-2887
E-mail:
Other service addresses can be obtained through the address indicated
above.
Manufacturer’s liability
We expressly point out that we can accept responsibility for any effects
on the safety, reliability, and performance of our medical products, if and
only if:
• any assembly, extensions, readjustments, modifications, or repairs are
performed by technically experienced, knowledgeable and trained
personnel and
• the medical products are used as set forth in the instructions for use.
It is possible to learn servicing the respective medical products through
appropriate courses of instruction offered by Aesculap. To arrange for such
a course, please contact Aesculap Technical Service (ATS).
To ensure that your Aesculap warranty remains valid, we recommend
using for repairs only Aesculap factory-tested spare parts. Spare parts, as
well as the relevant tools, can be ordered from Aesculap Technical Service
(ATS). Any unauthorized opening and/or alterations of the medical product
by third parties lead to the exclusion of our liability, as far as a fault is
attributable to such unauthorized opening and/or alteration of the
product. Aesculap cannot accept responsibility for the use of unsuitable
spare parts, tools or devices. After repairs, dropping, severe damage or
misuse, the product should be inspected by a qualified person.
This manual refers to the product as it was when the manual was
prepared. Technical changes, particularly of software, may be made at any
time.
General advisory
This service manual contains illustrations and explanations. Each
explanation covers the following: modes of operation, parts list, assembly,
disassembly, functionality test, and a tools/devices section (if applicable).
For further information about a specific product (e.g., troubleshooting list,
accessories) please refer to the relevant user instructions.
It is crucial to maintain extreme cleanliness during repairs of the products
described below. All products must undergo thorough function tests after
any repairs. If, in the course of a repair, a labeled component is replaced,
the labeling must be transferred to the spare part. Either the original label
should be used or the spare part bearing the relevant label should be
ordered. Any such order should include the relevant information.
Unless otherwise indicated, all medical products should be inspected
annually.
General information about motors and handpieces
It is critically important that ball bearings be inspected annually. Even the
slightest defect or dirt can lead to overheating during use, which would
render the product unusable. If for the repair of e.g. a motor a ball bearing
has to be replaced, it is best in most cases to also replace all other ball
bearings.
If it is necessary to heat up adhesive joints during disassembly, a hot-air
heat gun is the appropriate tool. The smallest possible nozzle should be
used so as to avoid damaging other components. For glue joints, make
certain that the surfaces to be joined are absolutely clean and free of
grease. We recommend using Loctite Rapid Cleaner 7063 for this purpose
(WS. no. 520001750). Only use adhesives listed in this manual. Follow the
glue manufacturer’s recommendations.
Lubricate components with the recommended Aesculap product only.
Following disassembly, clean all components thoroughly and inspect them
for any damage. Do not install any component of uncertain condition.
Products must always be test-run for several minutes after any repair.
Documentation
Art. no.
Instructions for use
TA009539