Recommended separation distances between portable and
mobile RF communication equipment and the Aeroneb Pro
The Aeroneb Pro nebulizer system is intended for use in the
electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Aeroneb Pro nebulizer
system can help prevent electromagnetic interference by maintaining a
minimum distance between port and mobile RF communications
equipment (transmitters) and the Aeroneb Pro nebulizer system as
recommended below, according to the maximum output power of the
communications equipment.
Rated maximum
output power of
transmitter
W
Separation distance according to frequency
of transmitter
m
150 kHz to
80 MHz
d
= [1.17] √P
80 MHz to
800 MHz
d
= [1.17] √P
800 MHz to
2.5 GHz
d
= [2.33] √P
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.75
1
1.17
1.17
2.33
10
3.70
3.70
7.36
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above,
the recommended separation distance d in metres (m) can be
estimated using the equation applicable to the frequency of
the transmitter, where P is the maximum output power rating of the
transmitter in watts (w) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the
higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations.
Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
Содержание Aeroneb Pro
Страница 1: ...Instruction Manual ...
Страница 2: ... ...
Страница 3: ...Aeroneb Professional Nebulizer System Instruction Manual ...
Страница 4: ... This page has been intentionally left blank ...
Страница 48: ...42 Aeroneb Professional Nebulizer System Instruction Manual This page has been intentionally left blank ...
Страница 49: ...Aeroneb Professional Nebulizer System Instruction Manual 43 Appendix 1 EMC Tables ...
Страница 57: ......