Professional Nebulizer System Instruction Manual
5
Warnings
To avoid damage to the nebulizer:
Prior to use, autoclave according to specified
directions and temperature given in the Cleaning,
disinfection and sterilization section of this
Instruction Manual only. Any deviation from
directions given in this Instruction Manual may
cause damage to the nebulizer unit and render
it inoperable.
Do not apply undue pressure to the domed aperture
plate in the center of the nebulizer.
Do not push out the Aerogen Vibronic
®
Aerosol
Generator.
Do not use a syringe with a needle to add
medication.
Do not use abrasive or sharp tools to clean the
nebulizer unit.
Inspect all parts before use, and do not use if any parts are
missing, cracked or damaged. In case of missing parts,
malfunction or damage, contact your Aeroneb Pro nebulizer
system sales representative.
Disconnect nebulizer unit from control module before cleaning.
Do not immerse or autoclave the control module or AC/DC
adapter.
Disassemble all parts before autoclaving.
Use only with components specified by Aerogen.
Do not use or store outside of specified environmental
conditions.