16
Aeroneb
®
Professional Nebulizer System Instruction Manual
Installation for use with a ventilator
1. For
adult breathing circuits
, connect the nebulizer unit
with adult T-piece into the inspiratory limb of the breathing
circuit before the patient Y (Figure 6).
For
pediatric breathing circuits
, connect the nebulizer
unit with pediatric T-piece into the inspiratory limb of the
breathing circuit before the patient Y (Figure 7).
For
neonatal breathing circuits
, connect the nebulizer
unit with the pediatric T-piece and the neonate adapters
approximately 30 cm (12 in.) back from the patient Y
(Figure 8). Alternatively connect the nebulizer with the
Neonate T-piece 30 cm (12 in.) back from the patient Y
(Figure 9).
WARNING:
Always maintain the nebulizer in a vertical orientation
(with the filler cap uppermost) while in the patient circuit
(Figures 6, 7, 8 and 9). This orientation prevents
condensate from blocking the nebulizer and ensures
proper nebulization. Always visually inspect the nebulizer
prior to placing in the ventilator circuit to assure that no
secretions are blocking the Aerogen Vibronic
®
Aerosol
Generator.
When removing the nebulizer unit from the patient circuit
always replace the T-piece plug to maintain circuit
pressure.
Condensate can collect and occlude ventilator circuits.
Always position ventilator circuits so that fluid condensate
drains away from the patient.
Always connect a bacteria filter to the expiratory inlet of
the ventilator. Otherwise the function of the expiratory
channel may be degraded.
Содержание Aeroneb Pro
Страница 1: ...Instruction Manual ...
Страница 2: ... ...
Страница 3: ...Aeroneb Professional Nebulizer System Instruction Manual ...
Страница 4: ... This page has been intentionally left blank ...
Страница 48: ...42 Aeroneb Professional Nebulizer System Instruction Manual This page has been intentionally left blank ...
Страница 49: ...Aeroneb Professional Nebulizer System Instruction Manual 43 Appendix 1 EMC Tables ...
Страница 57: ......