42
Simple Logger
®
II Series
6.2
Replacing the Fuse
(Model L111)
CAUTION:
Risk of electric shock. Disconnect all input(s) from the unit or
remove the clamp from any conductor before replacing the fuse.
•
With a flat-head screwdriver, twist the spring-loaded fuse holder
one quarter turn counter-clockwise.
•
The fuse holder will automatically eject.
•
Replace with a new 2A FA 250V 5x20mm fuse, then replace
holder.
6.3 Cleaning
CAUTION:
Risk of electric shock. Remove the clamp from any conductor
or disconnect all input(s) from the unit.
The body of the logger should be cleaned with a damp and soapy cloth.
Do not submerge the logger in water. Do not use solvent.
For the Model CL601, it is important to keep the jaw mating surfaces clean
at all times. Gently clean the clamp-on jaw mating surfaces with a soft
cloth and lightly oil them to prevent rust. Do not use water, solvents or
leave any residue.