50
Sentinel
™
Data Loggers
XRM Recording Mode
In the XRM mode, you only specify the starting date and time for the
recording.
The XRM mode operates much like Start/Stop mode with the exception
that the recording does not stop when memory becomes full. When this
happens every other sample is discarded, memory is compacted and
storage continues at 1/2 the storage rate. This mode of recording adjusts
the storage rate in order to keep the recording going. The instrument will
continue to record until the user stops it.
• Select
the
XRM
radio button.
• To specify and enable a recording, place a check in the
Schedule
Recording
check box (located under the Storage rate drop-down
menu).
Figure 5-26
Recording Start
You can specify a start date and time in one of two ways:
• Select
the
Now
button. Doing so will set the
Recording Start
date and time to the current date and time.
• Enter the start date and time in the associated edit boxes.