MiniFlex
®
Flexible AC Current Probe
3
1.1
International Electrical Symbols
This symbol signifies that the instrument is protected by double or reinforced
insulation.
This symbol on the instrument indicates a WARNING and that the operator must
refer to the user manual for instructions before operating the instrument. In this
manual, the symbol preceding instructions indicates that if the instructions are
not followed, bodily injury, installation/sample and product damage may result.
Risk of electric shock. The voltage at the parts marked with this symbol may
be dangerous.
This symbol refers to a type B current sensor. Do not apply around or remove
from HAZARDOUS LIVE conductors without additional protective means (de-
energizing the circuit or wearing protective clothing suitable for high voltage
work).
Battery
Compliance with the Low Voltage & Electromagnetic Compatibility European
directives (73/23/CEE & 89/336/CEE)
In the European Union, this product is subject to a separate collection system
for recycling electrical and electronic components In accordance with directive
WEEE 2002/96/EC
1.2
Definition of Measurement Categories
CAT II:
For measurements performed on circuits directly connected to
the electrical distribution system. Examples are measurements
on household appliances or portable tools.
CAT III:
For measurements performed in the building installation at
the distribution level such as on hardwired equipment in fixed
installation and circuit breakers.
CAT IV:
For measurements performed at the primary electrical supply
(<1000V) such as on primary overcurrent protection devices,
ripple control units, or meters.