AC Current Probe Model MA114
9
2. Turn the instrument ON by turning the rotary switch to any position. The
green ON indicator starts blinking.
3. Press on the yellow sensor opening button to open the flexible coil.
4. Place the open sensor around conductor to be measured (only one
conductor in the sensor at a time). Close the coil; an audible click
sounds as the coil is closed.
Use individual protection equipment when measuring an
uninsulated conductor carrying a high voltage.
To optimize measurement, center the conductor in the coil.
5. Set the switch to the range providing the best sensitivity and ensure the
red
OL
indicator is off. (
OL
indicates saturation of the electronics,
resulting in a measurement error.)