AC Current Probe Model MA114
3. SPECIFICATIONS
3.1. REFERENCE CONDITIONS
Quantity of influence
Reference values
Temperature
77 ± 9 °F (23 ± 5 °C)
Relative humidity
20 to 75 % RH
Measurement frequency range
30 to 440 Hz
Type of signal
sinusoidal
Instrument turn-on time
1 minute
External electric field
null
External DC magnetic field (earth's field)
< 40 A/m
External AC magnetic field
null
Position of the conductor in the coil
centered
Shape of the measurement coil
circular, no applied stress
Input impedance of the instrument connected
to the electronic unit
≥
1 M
Ω
3.2. ELECTRICAL CHARACTERISTICS
Parameter
Value
Range
3 A
30 A
300 A
3000 A
Specified measurement
range
0.5 to 3 A
2 to 30 A
5 to 300 A
50 to 3000 A
Output/input ratio (mV~/A~)
1000
100
10
1
Maximum peak factor
1.5 at I
N
(I nominal)
Intrinsic uncertainty
(% of output signal)
≤1%
+ 40 mV
≤1% + 4 mV
≤1.5% + 0.4 mV (I<10% I
N
)
≤1% + 0.4 mV (I≥10% I
N
)
≤1.5%+ 0.04 mV (I<10% I
N
)
≤1% + 0.04 mV (I≥10% I
N
)
Maximum offset voltage
0 mV
DC
Phase shift at 50 Hz
≤1° (0.5° typical)
Max. output voltage
-
4.5 Vpeak ≤ V ≤ + 4.5 Vpeak
Frequency response*
10 Hz to 10 kHz
10 Hz to 20 kHz
*Above 300 A
RMS
, see the graph in § 3.5.
The peak value is limited only by the electronics of the measurement
unit (red
OL
indicator on). The coil alone can briefly withstand higher
peak factors with no risk of overheating or damage.