6
Q
Place the selector switch on the
function :
- in forward direction the display gives the value of the junction
in V (resolution 1 mV)
- accuracy : ± 2% rdg ± 15 mV
- open circuit voltage : 1.5 V typical
- short circuit current : 0.4 mA
- in reverse direction the display indicates the open circuit
voltage, around 2 V.
Q
Freeze the value, if necessary, by pressing the HOLD button.
Q
Protection : idem Resistances
7- GENERAL SPECIFICATIONS
7-1
Dimensions and weight
Q
2.5" x 7" x 1.7" (64 x 177 x 42 mm)
Q
12.3 ounces (350 g)
7-2
Power supply
Q
Two batteries 1.5 V (type R6)
Q
Battery life : 300 h with R6 batteries
500 h with alkaline battery LR 6
Q
Low battery indication :
Note :
The voltage of the batteries is indicated approximately on the
diode function on open circuit.
7-3
Buzzer
Q
Continuous beep for the continuity test
Q
Intermittent beep each time the switch or buttons are pressed,
and for the overload indication.
7-4
Climatic conditions
Q
Temperature
- Use : 0°C to +50°C
- Storage: -20°C to +60°C
Q
Relative humidity
- Use :
≤
70 % RH
- Storage :
≤
90 % RH (up to 45°C)
Q
Altitude : use < 2000 m
7-5
Conformity with international standards
7-5-1 Electrical safety (IEC 1010-1)
Q
Double insulation :
Q
Installation category : III
Q
Degree of pollution : 2
Q
Assigned voltage : 600 V
Note :
This multimeter, of overvoltage category III satisfies the severe
requirements of reliability and availability corresponding to
industrial and domestic permanent installations (IEC 664-1).