5
V DC and AC
(1)
200 mV
2 V
20 V
200 V
600 V
Resolution
0.1 mV
1 mV
10 mV
100 mV
1 V
Impedance
3 M
Ω
Accuracy
on DC : ± 1 % reading ± 1 count
on AC : ± 1 % reading ± 5 counts
Permitted
600 V rms and 900 V peak
overload
(1) Use on frequency : 40 Hz to 500 Hz
Min. threshold on V AC : 25 mV from 40 Hz to 100 Hz and 300 mV from 100 Hz to 500 Hz
Min. threshold on V DC : 10 mV
4 - RESISTANCE
Never test a resistance on a live circuit
Q
Connect the leads to the multimeter and to the circuit or the
component to be tested.
Q
Place the selector switch on the ohmmeter function
Ω
Q
Automatic range selection : read the measured value
Q
Freeze the value, if necessary, by pressing the HOLD button.
Ω
Ω
Ω
Ω
Ω
200
Ω
2 k
Ω
20 k
Ω
200 k
Ω
2 M
Ω
20 M
Ω
Resolution
0.1
Ω
1
Ω
10
Ω
100
Ω
1 k
Ω
10 k
Ω
Accuracy
± 1.5 % rdg ± 1.5 % rdg
± 3 % rdg ± 5 counts
± 5 counts ± 3 counts
Voltage on
≤
0.5 V
open circuit
Protection
500 V rms and 750 V peak
5 - CONTINUITY BEEPER
Q
Place the selector switch on the
function
Q
Continuous audible beep emitted for a resistance R
≤
700
Ω
6 - DIODE TEST
Never test a diode on a live circuit
Q
Connect the leads to the multimeter and to the terminals of the
component to be tested. With this function, it is possible to test,
besides classic diodes, LED’s or any other semiconductor in which
the junction corresponds to a direct voltage less than 3 V.