message in the form Err XX.
•
Harmless
Errors
6
,
7
,
11
The message appears for approximately 1 second to inform
the user. Consider a repair if the error recurs.
• Errors
6
and
7
are always preceded by an automatic reset.
• During Error
11
, the tester automatically resets to the fac
tory default settings.
•
Correctable
Errors
5
,
14
,
15
,
18
,
19
,
30
,
31
,
32
and
33
These errors concern the current measurement function and
be used, but a repair will be necessary if the error persists.
• Error
18
indicates that the rechargeable battery in the
tester cannot be recharged. If this error occurs during
recharging, disconnect the tester from the charger and
follow the procedure described in the “fatal” errors.
• Error
19
can be corrected by clearing the entire memory of
the unit (see section 8.3).
• Errors
31
,
32
and
33
measurement setup for faults.
•
Fatal
Errors
0
,
1
,
2
,
3
,
8
,
12
,
13
,
15
,
16
,
18
(during battery recharge)
and
21
the unit and turn it on again. If the error persists, repair is
necessary.