AUTOMATIC
45
11.15
Defibrillation PADs
Category
ADULTS
CHILDREN
REF (Model)
SAV-C0846
SAV-C0016
Series
Cable and connector outside the envelope
Cable, connector and PADs into the envelope
)
Patient range
Adults age
>8 years
or weight
> 25Kg
Children age
1 - 8 years
or weight
< 25Kg
Intended use
Disposable
No. of shocks tolerated
50 shocks at 360J
Support material
Medical FOAM, thickness 1 mm
Conductive gel
Low impedance conductive adhesive gel
Total area (
for pad
)
148 cm ²
75 cm ²
Active area (
for pad
)
81 cm ²
31 cm ²
Conductive material
Metal foil
Connection
Anti-shock safety connector
Cable length
120 cm (normally)
11.16
Timing of Shock cycles
Charging time performance in accordance with 60601-2-4 (201.101)
Specific
Result
The maximum time between the beginning of the ECG rhythm analysis and the completion of
the charge at maximum energy
< 30 seconds
OK
The maximum time from turn on to completion of the charge at maximum energy
< 40 seconds
OK