AUTOMATIC
10
CAUTION
Do not allow defibrillation electrodes to touch or come into contact with ECG electrodes, swabs, transdermal patches, etc. Failure to do
so may result in creation of electric arcs and burns to the patient during defibrillation, and even current leakage.
Position the defibrillation PADS as indicated in this user manual and indicated on the package.
Do not use defibrillation PADs if the gel has been detached from the support or is torn, split or dry.
If damage has been detected, do not operate the
Saver One
under any circumstances.
Before using the device, remove metal objects from the patient's body (including necklaces or bracelets, etc.)
Do not use defibrillation PADs other than those supplied by the manufacturer. Otherwise the defibrillator may make false
interpretations.
Do not use defibrillation PADs if they are damaged, even partially.
Do not touch the patient or PADs during heart rhythm analysis.
Moving or transporting the patient during the cardiac rhythm analysis performed by the device can lead to an incorrect or not timely
diagnosis. During the heart rhythm analysis phase, minimize the movements. If the device is used in an ambulance in motion, stop the
vehicle and start again only after having delivered the shock.
In order to use the
Saver One
, you must have completed a training course for basic or advanced cardio-pulmonary resuscitation with
the use of a defibrillator (BLS-D or ALS-D course)
Avoid the use of adult defibrillation PADs (SAV-C0846) on children (ages 1-8 years or 8-25kg).
Before applying the defibrillation PADS, if necessary, dry the patient's chest and remove unwanted hair.
Do not subject
Saver One
, its accessories, its parts to falls and / or strong impacts
Do not use damaged accessories and / or parts, otherwise the device may malfunction.
Use only original accessories and / or spare parts.
Avoid excessively aggressive handling of the device of its accessories or parts in order to avoid possible damage. Inspect the entire
system periodically.
Carry out the sanitation operations of the device in compliance with the standards indicated in paragraph 10.3 and always make sure that
the device is switched off with the battery removed and PADs disconnected.
Defibrillation PADs are disposable, to be used only on one patient. Do not reuse defibrillation pads; discard after use and replace with a
new pair.
Defibrillation PADs are not sterile or sterilizable.
Intense or prolonged administration of cardiopulmonary resuscitation with defibrillation electrodes applied to the patient can damage the
electrodes. Replace them if they are damaged due to use or handling.
Improper maintenance can damage the
Saver One
or cause it to malfunction. Follow the instructions in this user manual.
Use original non-rechargeable Li-SOCl2 (SAV-C0903) batteries from A.M.I. Italia S.r.l. before the indicated expiration date.
Recharge the rechargeable Li-ion battery (SAV-C0011) at least once every 4 months ensure its perfect function and extend its life.
The Li-ion rechargeable batteries ACC model (SAV-C0011) must be charged using only the SAV-C0012 battery charger from A.M.I.
Italia S.r.l. otherwise the batteries could be damaged
Remove the batteries from the device only if it has been turned off for at least 5 seconds. Otherwise the device and the battery can be
seriously damaged.
The
Saver One
, its parts and accessories are not sterile or sterilizable
Do not expose the
Saver One
, its parts or accessories to direct light or high temperatures
The Battery Charger (SAV-C0012) must only be used with the Meanwell power supply model GS40A15-P1J (SAV-C0013) supplied by
A.M.I. Italia S.r.l. The use of different power supplies could compromise the correct functioning of the battery charger and damage the
ACC rechargeable batteries (SAV-C0011)
In order to safeguard the battery life (SAV-C0903) and guarantee automatic daily tests, after installing it, it is advisable to not remove
the battery (SAV-C0903) unless it is to be replaced. The removal of the battery and the subsequent insertion involves a complete test of
the AED which considerably consumes its capacity. Furthermore, if the battery is not properly attached it could be damaged.
2.3
Indications of
DISPOSAL
The
Saver One,
its parts and accessories must not be disposed of with other household waste within the European community. To prevent
possible damage to the environment or human health caused by incorrect waste disposal, recycle this product responsibly also to promote
sustainable use of resources. To dispose of the used product, use the appropriate waste collection services or return it to the local distributor. In
this way it will be possible to recycle safely for the environment
Содержание Saver One
Страница 1: ...User Manual External AUTOMATIC Defibrillator for public access Rev 12 2...
Страница 2: ...AUTOMATIC 2 QUICK USE GUIDE...
Страница 50: ...AUTOMATIC 50 14 Certifications 14 1 EC Certificate...
Страница 51: ...AUTOMATIC 51...
Страница 52: ...AUTOMATIC 52 14 2 IMQ Brand...
Страница 55: ...AUTOMATIC 55...