4
If you enter a width of zero, the unit will change to CW mode. Entering a non zero width
will convert the unit into sweep mode. The display formats are different for these two modes.
Another way to toggle these modes is explained in paragraph 3.8.4.1 .
Due to synthesizer limitations, the sweep widths must be certain values and the VIA
Bravo MRI adjusts your entry to an available sweep width. For example if you enter a 1.000
MHz wide sweep, the VIA Bravo MRI sets the sweep width to 1.000 MHz if the unit is plotting
100 points, but will set the sweep width to 1.600 MHz if the graph is set to 80 points. The VIA
Bravo MRI flashes a brief warning if it changes the sweep width from the number you
entered. The x axis legend displays the lower and upper sweep frequencies of the chosen
sweep bandwidth.
Sweep width may also be set using the WIDTH keys. Pressing one of these keys will
increase or decrease the sweep width by a factor of 2. The WIDTH key may be repeatedly
pressed for larger changes (i.e. press the width key 3 times to get a sweep that is 8 times
larger).
Some combinations of center frequency and sweep width will cause the graph to be
truncated. The sweep truncates at the VIA Bravo MRI’s frequency limits. Truncation can
also occur using the EXAM/PLOT mode, see paragraphs 3.4 for details on this.
If you make an error while entering the sweep width, you can continue to enter digits (use
zeros) until an illegal width (too wide) is entered. Eventually the display resets, allowing you
to start a new sweep entry.
You are allowed to interrupt calibration cycles when you enter a width command. A new
calibration cycle will start after you enter your new width.
3.3 Frequency Step Size
The frequency step size determines the change in center frequency when the FREQ key
is pressed. The center frequency may be adjusted up or down by using the appropriate
FREQ key.
Setting the frequency step size is similar to center frequency entry, except use FREQ
instead of ENTER.
The frequency step sizes are limited by the sweep parameters. If you enter an
unavailable step size, the VIA Bravo flashes a brief warning and sets the step to a valid
value. The higher the center frequency, the wider the available sweep width.
Exit any menus that you may be in then press the first digit of your desired step
frequency. The frequency entry screen pops up. Finish entering the step frequency. Note
that you may need to add zeroes to get your entry to align properly with the decimal point.
When the correct number is ready, press either FREQ key. The unit should start plotting with
the new center frequency. Any future presses of the FREQ keys will increment the center
frequency by this new step frequency.
If you make an error while entering the step size, you can continue to enter digits until an
illegal step (too big) is entered. When you press the FREQ or ENTER keys, the display
resets, allowing you to start a new step size entry.