32
no effect on the unit’s setup. These read only values will be sent to the
terminal in response to a S000 or Mxx command.
Setup Data Table
Characters
Received
Range of x…x
Meaning
Notes
A
S002
Header denotes unit
setup info follows
Toss this out
Fx…x
100,000 to
70,000,000
Center frequency in
Hz
Wx…x
0 to 25.6MHz or
32MHz
Sweep width in Hz
Dxxx
101 to 104
Data format
Overrides plot type if in
conflict.
x=101; R and X
x=102; Z and A
x=103; V and L
x=104; M and D
Ax…x
0 to 1
Instrument mode
0=VIA, 1=SWR
Ax…x
24
Memmax
Maximum memory
locations
Ax…x
0 to 200
CW index
Factory use
*
Ax…x
0 to 50
Lower plot index
Equal to 0 except when
the sweep would try to
extend below the
minimum calibrated freq
*
Ax…x
40 to 100
Upper plot index
Usually the same as
points in a sweep except
when sweep extends
beyond maximum cal. freq
*
Ax…x
0 to 1
Auto power shutoff
0-disabled, 1-enabled
Ax…x
0 to 3
Calibration Mode
0,1 =sweep; 2,3 =CW
1,3 =Cable null mode
Ax…x
0 to 255
Backlight Intensity&
Contrast
low nibble is intensity
f =off,e=bright,0=dim
High nibble for contrast
0=light, f=dark
Ax…x
0 to 4
Backlight Timer
0=off, …,4=continuous
Ax…x
1, 3, 5
Horizontal grids
Ax…x
0 to 1
Big Freq Display
0=off; 1=on
Ax…x
0 to 3
Audio Volume
0=soft;…,3=loud
Ax…x
0 to 2
Audio Mode
0=off; 1=on;
2=audio readout
Ax…x
0 to 8
Left plot type
0=Z;1=A;2=R;3=X;
4=RCM;5=RCA;
6=none;7=VSR;8=RL