background image

21 

 

Broadband

 is a calibration that is done once, 

using the Site Analyzer’s entire frequency range 

100KHz to 450MHz and saved for use with any shorter frequency ranges selected for use. The 
advantage of using Broadband Calibration will be that recalibrating the instrument for every change of 
frequency will not be needed. It can also include any combination of test leads or adapters, effectively 
removing them from the measurements. However, that same test lead / adapter combination MUST 
be used with all SWR or RL tests performed when Broadband Calibration is being used. The 
disadvantage of using Broadband Calibration is that it will not be as accurate as 

Frequency Specific 

Calibration.

 

Frequency Specific Calibration

 is performed with any frequency selection or adjustment of the 

Start, Stop, Center Frequency or Span. The advantage of Frequency Specific Calibration is it 
provides the maximum accuracy possible by the MRI-3000. The disadvantage is that it requires a 
recalibration process with any frequency change. Frequency Specific Calibration can be done with 
any combination of test leads or adapters, allowing them to be effectively removed from the 
measurements.  When prompted, apply the terminato

rs (Open, Short and 50Ω Load) to the far-end of 

the test lead (if you're using those), or directly to 

the instrument’s N connector. If a test lead and/or 

adaptor is used, they must remain attached during any measurements at the calibrated frequency 
range. 

 

Broadband Calibration Process 

Preliminary: To perform this process, first decide if you will be using an adapter, test lead or both. 
Remember if you do, that same test lead and/or adapter MUST be used with all measurements 
employing the Broadband Calibration. 

Second, make sure you have the correct terminators OPEN, SHORT and LOAD with correct gender 
to fit either the far-

end of the test lead or the instrument’s N connector. 

Steps: 

1. 

Press the SWR/RL key to select either measurement type. 

2. 

Press Mode key and select MORE OPTIONS. Press ENTER. 

3. 

From the OPTIONS menu, select BROADBAND CAL, press ENTER. 

4. 

First prompt will be to attach: OPEN.  As said previously the open N connector, adapter, or 

test lead will suffice. Using an 

OPEN Terminator is optional at user’s discretion. 

5. 

Press ENTER to continue. 

6. 

Second prompt will be to attach the SHORT Terminator. 

7. 

When attached press ENTER to continue 

8. 

The l

ast prompt will be to attach the LOAD. Normally this will be a 50 Ω load supplied with the 

instrument or purchased locally. 

9. 

When attached press ENTER to continue and end the Broadband calibration process. The MRI-

3000 will save the Broadband calibration, set the calibration mode to BROADBAND, and return 
to the SWR/RL Measurement screen. You now select the frequency (via the FREQ key), and 
attach the device to be tested. The SWR/RL key switches between SWR and Return Loss.

 

 

Frequency Specific Calibration Process 

Preliminary: To perform this process, first decide if you will be using an adapter or test lead or both. 
Remember if you do, that same test lead and/or adapter MUST be used with all measurements 
employing the Frequency Specific Calibration. 

Second, make sure you have the correct terminators OPEN, SHORT and LOAD with correct gender 
to fit either the far-

end of the test lead, adapter or the instrument’s N connector. 

Steps: 

1. 

If not already in Frequency Specific Calibration mode, press Mode Menu key. Otherwise 

Содержание Bravo MRI 3000

Страница 1: ...Operation Manual Liberator Series Bravo MRI 3000 Analyzer Testing Made Simple _ Technology Inc...

Страница 2: ...his document and all software or firmware designed by AEA Technology Inc is copyrighted and may not be copied or altered in any way without the written consent from AEA Technology Inc The AEA Logos ar...

Страница 3: ...nd 5 feet 1 5 meter with a 3T Coil The MRI Analyzer when in its Belt Case is splash resistant and designed to work for relatively long periods in rain depending on intensity It is not designed to be i...

Страница 4: ...of the semiconductor devices used in this instrument are susceptible to damage by static discharge Static charges are generated in numerous ways such as simple contact separation of materials and nor...

Страница 5: ...ad menus and measurement screens 2 Provides user instructions in more detail than the Quick Start Guide including a section on Measurement Information and Helpful Tips 3 Provides specifications troubl...

Страница 6: ...tion 1 General 1 Instrument Front Panel Layout 2 Screen Layouts 2 General 2 VNA Wide Format 2 VNA Narrow Format 4 SWR RL Wide Format 4 SWR RL Narrow Format 5 S21 Screen format 6 Keypad Operations 6 Po...

Страница 7: ...ss Screen 18 CW Continuous Wave Screen 18 Cable Null Screens 18 Calibration Sequence Screens 19 MRI 3000 VNA Measurements 19 General Information 19 S11 VNA SCREEN 19 S21 VNA Screen Two Port Model only...

Страница 8: ...nce and Troubleshooting Guides 37 Limited Warranty 37 Maintenance 37 Battery Replacement Instructions 37 Cleaning 38 Instrument Troubleshooting Guide 39 Factory Reset 39 Used to recover any program fa...

Страница 9: ...cement for network analyzer applications that measure S11 4 Find resonant frequency and response curve 5 CW signal generator 6 Two plots may be simultaneously viewed on the same graph The MRI 3000 com...

Страница 10: ...y plots on a chart This graph will display a different number of points depending on the width selected The sidebar section to the left of the chart numerically indicates measurements at center freque...

Страница 11: ...ery sweep It should be noted that all of the above elements are common to both the VNA and SWR mode wide format screens As previously mentioned the MRI 3000 s screen is composed of the chart area and...

Страница 12: ...tor for the value of Q for the circuit under test In VNA mode the Q factor is displayed only when narrow chart format is selected shown here and only if the impedance magnitude Z MAG is selected for e...

Страница 13: ...eft scale displayed and selected by pressing the SWR RL key Due to there being only two graph possibilities SWR RL on this screen the top indicator will always display the SWR at center frequency the...

Страница 14: ...ault settings There are two different power down sequences as described below the Power ON POWER ON Press the ON OFF key for one second A splash screen with AEA Technology s logo and Version informati...

Страница 15: ...a saved setting or set desired settings Menu Keys These keys bring up the Meter Settings Options settings and Memory Operations menus respectively For details please refer to the Menus section on page...

Страница 16: ...ted and measurements appear in graphic plots Continuous Wave mode stops the Sweep mode and only emits a continuous wave at the Center Frequency This changes the graphic plot to a list of tests showing...

Страница 17: ...ly functions with one or both cursors enabled In SWR and RL pressing Auto Tune will cause the selected cursor to seek to the LOWEST point on the graph In VNA Mode the cursor will seek to the highest i...

Страница 18: ...or setting CAUTION There is NO UNDELETE Once a test result or setup is erased it is permanently deleted The alpha numeric keypad operates like those on smaller cell phones Press a selected key once f...

Страница 19: ...om for each item on the list 3 Use the ENTER key as indicated to call a sub menu of selections Navigation in sub menus works the same 4 When selections are completed use the ESC key to save and exit t...

Страница 20: ...isible KEYPAD BEEP ON or OFF Press to make a selection UNITS FEET or METERS Press to make a selection BATTERY MENU Press ENTER to open sub menu Selecting the BATTERY MENU item from the Meter menu will...

Страница 21: ...ESC to exit to Measurement Screen Mode Menu The Mode Key of the MRI 3000 will open the OPTIONS screen This allows you to change the various measurement options and settings for the instrument Use UP...

Страница 22: ...on will initiate the broad banded calibration cycle This calibration is stored in the unit and is valid anywhere in the units operating range The Broadband calibration may be performed with or without...

Страница 23: ...or C to be calculated correctly the user must specify which circuit model to use SYSTEM Z0 Allows the setting of the MRI 3000 s reference impedance Z0 to something other than 50 which is the default...

Страница 24: ...cy Bandwidth Span Screens Frequency Specific Calibration Screens Calls a series of Prompt screens to attach the Open Short and Load Terminators followed by using ENTER to process calibration data for...

Страница 25: ...enable the second CRSR B Cursors can be moved to the desired frequency via the LEFT RIGHT arrow key If both cursors are enabled pressing the CRSR key will toggle the cursor that is currently selected...

Страница 26: ...ewing of any changes during DUT tuning being performed To exit press the Sweep CW key again to return to the previous testing mode and the swept measurements will resume Cable Null Screens SWR RLkeysw...

Страница 27: ...f the measurement test results which may or may not be desired Calibration Sequence Screens Pressing ENTER after the Load is attached completes the calibration process for single port measurements and...

Страница 28: ...Xs reactance With the desired parameter selected press SCALE to select the scale to best view the traces With the cursor enabled via the CRSR key the measurements at the cursor location are displayed...

Страница 29: ...Remember if you do that same test lead and or adapter MUST be used with all measurements employing the Broadband Calibration Second make sure you have the correct terminators OPEN SHORT and LOAD with...

Страница 30: ...the OPEN cal standard will yield the most consistent results 7 Press ENTER to continue 8 Second prompt will be to attach the SHORT Terminator 9 When attached press ENTER to continue 10 Last prompt wil...

Страница 31: ...ips 1 When taking SWR measurements on a new system measure the components separately first and document each item s SWR Then measure the antenna system as a whole and document that measurement This wi...

Страница 32: ...the usable frequency range clearly between the Cursors from low to high frequency SWR and Return Loss Measurement Steps SWR or Return Loss Measuring 1 Connect the test lead and any adapter if they ar...

Страница 33: ...tinent user data downloads traces etc will be stored there Running MRI PC Vision Start the MRI PC Vision Application by either clicking on the icon on your desktop or by selecting it from the start me...

Страница 34: ...this select COM Port Mgr Com Port from the dropdown menu at the top to open this dialog box Please Note You must ONLY change the COM port number DO NOT adjust any of the other settings Click on the dr...

Страница 35: ...cle of the instrument AFTER you ve set the frequency range you want to test Please note that this menu is for compatibility with earlier versions the S11 Calibrate button will do the same thing Basic...

Страница 36: ...tional OPEN standard to the MRI 3000 or attached cable and press the Ready button for step one The MRI PC Vision Program will assume control over the instrument and begin the calibration process When...

Страница 37: ...29 if calibration has not already been completed you will be prompted to do so and the measured results of the UUT unit under test will be drawn on the graphs as shown here...

Страница 38: ...ency scale along the bottom in MHz and the vertical scale control top left While the frequency scale is fixed it s determined by the frequency range you chose when you calibrated the instrument you ca...

Страница 39: ...est viewing Smith Chart The Smith Chart display allows you to easily visualize the UUT s S Parameter characteristics Unlike the other displays in this application you move the cursors using the slide...

Страница 40: ...ote that the cursor is at a point on the trace where the value in this case the SWR is at or below the limit line Note also that the limit line is GREEN If the cursor is at a place where the trace is...

Страница 41: ...ming Your Chart Each chart displayed on the screen has a default header describing the measurement displayed The default heading for the SWR chart is SWR Plot for example Any chart can be renamed by s...

Страница 42: ...up a file save dialog box Select the format to save your trace in give it a name and click SAVE It is advisable to save your traces as PCV format first You can later save individual charts in a standa...

Страница 43: ...s in your instrument s file system The sys files belong to the system and should not be downloaded or deleted unless instructed to do so by support personnel The RTL SWR and VNA files are Return Loss...

Страница 44: ...nstrument From the file management window shown earlier select the file s you wish to delete and click Delete CAUTION there is no UNDELETE function all saved test results and setups deletes are perman...

Страница 45: ...the instrument other than the battery compartment as this can void the warranty If the instrument is not performing or charging correctly refer to Troubleshooting Guides at the end of this section or...

Страница 46: ...t see recommended cleaners listed below sprayed on the cloth first Rinse with a clean soft damp cloth moistened in water only Do NOT spray detergents or water directly on the instrument Avoid using so...

Страница 47: ...ructions Instrument will not turn off Firmware has faulted Power down with 10 second hold of ON OFF key This will load PRIOR good settings NOT current settings See page 5 for more help and Soft Reset...

Страница 48: ...olution 1 Total Z Range 0 to 2K Resolution 1 Z Angle Range 180 Resolution 1 General Power requirement 12 18 VDC 650 mA min or 8 AA Batteries AC Adapter Included 110 240 VAC 50 60Hz 0 4A Output 15 VDC...

Страница 49: ...i B 0070 1208 USB extension cord 15 ft 0070 1210 Belt Case 5001 1002 CD includes Operating Manual Training PPT and MRI PC Vision software 6055 1220 Optional Accessories Part Number N to SMA Adapters a...

Страница 50: ...42 User s Notes...

Страница 51: ...43 5933 Sea Lion Place Suite 112 Carlsbad California 92010 Tel 1 800 258 7805 or 1 760 931 8979 Fax 1 760 931 8969 www aeatechnology com...