background image

23 

 

SWR (Standing Wave Ratio a.k.a. Voltage Standing Wave Ratio-VSWR) 

SWR is a measurement of impedance matching of loads to the characteristic impedance of a 
transmission line or waveguide. Impedance mismatches result in standing waves along the 
transmission line.  SWR is defined as the partial standing waves’ amplitude at an antinode 
(maximum) to the amplitude at a node (minimum) along the transmission path. 

It is often thought of as the maximum and minimum AC voltages along the transmission line lending 
to the term VSWR (Voltage Standing Wave Ratio). For example: with a VSWR = 1.3:1 an AC voltage 
due to standing waves along the transmission line is reaching a peak value 1.3 times that of the 
minimum AC voltage along that line. Hence, SWR can be defined as the ratio of the maximum 
amplitude to minimum amplitude of the transmission lines currents, electric field strength, or the 
magnetic field strength. Neglecting transmission line loss these ratios are identical. 

In general, the lower the SWR the better quality of the transmission line, component, antenna, or 
antenna system under test. Each component of a system will have SWR defined for that item at its 
designed frequency range. When an antenna system is designed the engineer will compute what the 
maximum SWR should be at center frequency for that particular system. When a new system is 
installed the SWR reading should meet or be low

er than maximum and be documented for “As Built” 

or “Base Line” measurements. Later measurements can then be compared to the As Built 
measurement to see if anything has changed. 

Tips: 

1. 

When taking SWR measurements on a new system, measure the components separately first 
and document each item’s SWR. Then measure the antenna system as a whole and document 
that measurement. This will help later when measuring a faulty system to identify the component 
that changed. 

2. 

When measuring a transmission line or line component in the line, be sure to attach an 
impedance matched terminator at the far-end. An un-terminated line or component will show a 
high SWR. Antenna’s are a termination in themselves 

3. 

Be  sure  about  the  line  or  component’s  center  frequency  for  which  it  was  designed  is  correctly 
entered  in  the  MRI-3000. 

This  will  center  the  SWR’s  dip  at  or  close  to  the  center  of  the 

measurement screen. 

4. 

If the lowest SWR reading is not on center frequency be sure to move the cursor to the 
planned use center frequency and read the SWR at that frequency. The lowest point in the 
waveform may meet system requirements, but the reading at the planned usage frequency is 
what counts and that should also be under the maximum design specifications. 

 

5. 

If you are probing an indeterminant or unknown system, it is recommended to use the MRI-
3000 with a Broadband Calibration starting with a wide frequency sweep. Gradually change 
the center frequency and span to locate the frequency(ies) of interest.  You can then switch to 
a frequency specific calibration to make more precise measurements if desired.

 

 

Return Loss 

Return Loss is the loss of power in a transmitted signal due to impedance mismatches or 
discontinuities in the transmission line, line components or antenna causing reflected power back to 
the transmission’s source. It’s expressed in dB as defined by the equation shown below: 

 

 

 

RL(dB)=Return Loss, Pi = incident power, and Pr = Reflected power 

Содержание Bravo MRI 3000

Страница 1: ...Operation Manual Liberator Series Bravo MRI 3000 Analyzer Testing Made Simple _ Technology Inc...

Страница 2: ...his document and all software or firmware designed by AEA Technology Inc is copyrighted and may not be copied or altered in any way without the written consent from AEA Technology Inc The AEA Logos ar...

Страница 3: ...nd 5 feet 1 5 meter with a 3T Coil The MRI Analyzer when in its Belt Case is splash resistant and designed to work for relatively long periods in rain depending on intensity It is not designed to be i...

Страница 4: ...of the semiconductor devices used in this instrument are susceptible to damage by static discharge Static charges are generated in numerous ways such as simple contact separation of materials and nor...

Страница 5: ...ad menus and measurement screens 2 Provides user instructions in more detail than the Quick Start Guide including a section on Measurement Information and Helpful Tips 3 Provides specifications troubl...

Страница 6: ...tion 1 General 1 Instrument Front Panel Layout 2 Screen Layouts 2 General 2 VNA Wide Format 2 VNA Narrow Format 4 SWR RL Wide Format 4 SWR RL Narrow Format 5 S21 Screen format 6 Keypad Operations 6 Po...

Страница 7: ...ss Screen 18 CW Continuous Wave Screen 18 Cable Null Screens 18 Calibration Sequence Screens 19 MRI 3000 VNA Measurements 19 General Information 19 S11 VNA SCREEN 19 S21 VNA Screen Two Port Model only...

Страница 8: ...nce and Troubleshooting Guides 37 Limited Warranty 37 Maintenance 37 Battery Replacement Instructions 37 Cleaning 38 Instrument Troubleshooting Guide 39 Factory Reset 39 Used to recover any program fa...

Страница 9: ...cement for network analyzer applications that measure S11 4 Find resonant frequency and response curve 5 CW signal generator 6 Two plots may be simultaneously viewed on the same graph The MRI 3000 com...

Страница 10: ...y plots on a chart This graph will display a different number of points depending on the width selected The sidebar section to the left of the chart numerically indicates measurements at center freque...

Страница 11: ...ery sweep It should be noted that all of the above elements are common to both the VNA and SWR mode wide format screens As previously mentioned the MRI 3000 s screen is composed of the chart area and...

Страница 12: ...tor for the value of Q for the circuit under test In VNA mode the Q factor is displayed only when narrow chart format is selected shown here and only if the impedance magnitude Z MAG is selected for e...

Страница 13: ...eft scale displayed and selected by pressing the SWR RL key Due to there being only two graph possibilities SWR RL on this screen the top indicator will always display the SWR at center frequency the...

Страница 14: ...ault settings There are two different power down sequences as described below the Power ON POWER ON Press the ON OFF key for one second A splash screen with AEA Technology s logo and Version informati...

Страница 15: ...a saved setting or set desired settings Menu Keys These keys bring up the Meter Settings Options settings and Memory Operations menus respectively For details please refer to the Menus section on page...

Страница 16: ...ted and measurements appear in graphic plots Continuous Wave mode stops the Sweep mode and only emits a continuous wave at the Center Frequency This changes the graphic plot to a list of tests showing...

Страница 17: ...ly functions with one or both cursors enabled In SWR and RL pressing Auto Tune will cause the selected cursor to seek to the LOWEST point on the graph In VNA Mode the cursor will seek to the highest i...

Страница 18: ...or setting CAUTION There is NO UNDELETE Once a test result or setup is erased it is permanently deleted The alpha numeric keypad operates like those on smaller cell phones Press a selected key once f...

Страница 19: ...om for each item on the list 3 Use the ENTER key as indicated to call a sub menu of selections Navigation in sub menus works the same 4 When selections are completed use the ESC key to save and exit t...

Страница 20: ...isible KEYPAD BEEP ON or OFF Press to make a selection UNITS FEET or METERS Press to make a selection BATTERY MENU Press ENTER to open sub menu Selecting the BATTERY MENU item from the Meter menu will...

Страница 21: ...ESC to exit to Measurement Screen Mode Menu The Mode Key of the MRI 3000 will open the OPTIONS screen This allows you to change the various measurement options and settings for the instrument Use UP...

Страница 22: ...on will initiate the broad banded calibration cycle This calibration is stored in the unit and is valid anywhere in the units operating range The Broadband calibration may be performed with or without...

Страница 23: ...or C to be calculated correctly the user must specify which circuit model to use SYSTEM Z0 Allows the setting of the MRI 3000 s reference impedance Z0 to something other than 50 which is the default...

Страница 24: ...cy Bandwidth Span Screens Frequency Specific Calibration Screens Calls a series of Prompt screens to attach the Open Short and Load Terminators followed by using ENTER to process calibration data for...

Страница 25: ...enable the second CRSR B Cursors can be moved to the desired frequency via the LEFT RIGHT arrow key If both cursors are enabled pressing the CRSR key will toggle the cursor that is currently selected...

Страница 26: ...ewing of any changes during DUT tuning being performed To exit press the Sweep CW key again to return to the previous testing mode and the swept measurements will resume Cable Null Screens SWR RLkeysw...

Страница 27: ...f the measurement test results which may or may not be desired Calibration Sequence Screens Pressing ENTER after the Load is attached completes the calibration process for single port measurements and...

Страница 28: ...Xs reactance With the desired parameter selected press SCALE to select the scale to best view the traces With the cursor enabled via the CRSR key the measurements at the cursor location are displayed...

Страница 29: ...Remember if you do that same test lead and or adapter MUST be used with all measurements employing the Broadband Calibration Second make sure you have the correct terminators OPEN SHORT and LOAD with...

Страница 30: ...the OPEN cal standard will yield the most consistent results 7 Press ENTER to continue 8 Second prompt will be to attach the SHORT Terminator 9 When attached press ENTER to continue 10 Last prompt wil...

Страница 31: ...ips 1 When taking SWR measurements on a new system measure the components separately first and document each item s SWR Then measure the antenna system as a whole and document that measurement This wi...

Страница 32: ...the usable frequency range clearly between the Cursors from low to high frequency SWR and Return Loss Measurement Steps SWR or Return Loss Measuring 1 Connect the test lead and any adapter if they ar...

Страница 33: ...tinent user data downloads traces etc will be stored there Running MRI PC Vision Start the MRI PC Vision Application by either clicking on the icon on your desktop or by selecting it from the start me...

Страница 34: ...this select COM Port Mgr Com Port from the dropdown menu at the top to open this dialog box Please Note You must ONLY change the COM port number DO NOT adjust any of the other settings Click on the dr...

Страница 35: ...cle of the instrument AFTER you ve set the frequency range you want to test Please note that this menu is for compatibility with earlier versions the S11 Calibrate button will do the same thing Basic...

Страница 36: ...tional OPEN standard to the MRI 3000 or attached cable and press the Ready button for step one The MRI PC Vision Program will assume control over the instrument and begin the calibration process When...

Страница 37: ...29 if calibration has not already been completed you will be prompted to do so and the measured results of the UUT unit under test will be drawn on the graphs as shown here...

Страница 38: ...ency scale along the bottom in MHz and the vertical scale control top left While the frequency scale is fixed it s determined by the frequency range you chose when you calibrated the instrument you ca...

Страница 39: ...est viewing Smith Chart The Smith Chart display allows you to easily visualize the UUT s S Parameter characteristics Unlike the other displays in this application you move the cursors using the slide...

Страница 40: ...ote that the cursor is at a point on the trace where the value in this case the SWR is at or below the limit line Note also that the limit line is GREEN If the cursor is at a place where the trace is...

Страница 41: ...ming Your Chart Each chart displayed on the screen has a default header describing the measurement displayed The default heading for the SWR chart is SWR Plot for example Any chart can be renamed by s...

Страница 42: ...up a file save dialog box Select the format to save your trace in give it a name and click SAVE It is advisable to save your traces as PCV format first You can later save individual charts in a standa...

Страница 43: ...s in your instrument s file system The sys files belong to the system and should not be downloaded or deleted unless instructed to do so by support personnel The RTL SWR and VNA files are Return Loss...

Страница 44: ...nstrument From the file management window shown earlier select the file s you wish to delete and click Delete CAUTION there is no UNDELETE function all saved test results and setups deletes are perman...

Страница 45: ...the instrument other than the battery compartment as this can void the warranty If the instrument is not performing or charging correctly refer to Troubleshooting Guides at the end of this section or...

Страница 46: ...t see recommended cleaners listed below sprayed on the cloth first Rinse with a clean soft damp cloth moistened in water only Do NOT spray detergents or water directly on the instrument Avoid using so...

Страница 47: ...ructions Instrument will not turn off Firmware has faulted Power down with 10 second hold of ON OFF key This will load PRIOR good settings NOT current settings See page 5 for more help and Soft Reset...

Страница 48: ...olution 1 Total Z Range 0 to 2K Resolution 1 Z Angle Range 180 Resolution 1 General Power requirement 12 18 VDC 650 mA min or 8 AA Batteries AC Adapter Included 110 240 VAC 50 60Hz 0 4A Output 15 VDC...

Страница 49: ...i B 0070 1208 USB extension cord 15 ft 0070 1210 Belt Case 5001 1002 CD includes Operating Manual Training PPT and MRI PC Vision software 6055 1220 Optional Accessories Part Number N to SMA Adapters a...

Страница 50: ...42 User s Notes...

Страница 51: ...43 5933 Sea Lion Place Suite 112 Carlsbad California 92010 Tel 1 800 258 7805 or 1 760 931 8979 Fax 1 760 931 8969 www aeatechnology com...

Отзывы: