![ADWA AD 630 Скачать руководство пользователя страница 38](http://html1.mh-extra.com/html/adwa/ad-630/ad-630_user-manual_2859272038.webp)
38
PROBE AND MEMBRANE MAINTENANCE
For a top performance probe, it is recommended to re-
place the membrane every 2 months and the electrolyte
once a month.
Proceed as follows:
• Unscrew the membrane by
turning it counterclockwise.
• Rinse the supplied spare
membrane with some electro-
lyte solution while shaking it
gently. Refill with clean elec-
trolyte.
• Gently tap the membrane over
a surface to ensure that no air
bubbles remain trapped.
• With the sensor facing down
screw the cap clockwise to
the end of the threads. Some
electrolyte will overflow.
If any deposit scales the sensors, gently brush the sensor
surface with the supplied scouring pad, while paying
attention to not damage the plastic body.
Содержание AD 630
Страница 8: ...8 FRONT AND TOP PANELS...