Note: If you are performing a slope calibration only,
press CAL and then use the ARROW keys to
select the 100% DO calibration point.
• Dry the probe tip and allow a few seconds for the
reading to stabilize. The hourglass symbol will blink
until the reading is stable.
• When the reading is stable, CFM starts blinking. Press
CFM to confirm the 100% calibration point.
• The instrument stores the slope calibration data and
returns to measurement mode.
CUSTOM CALIBRATION
AD630 allows to perform a 1 point calibration (slope
only) using a custom solution with a known DO value
greater than 20%.
Preparation
• Pour small quantity of the custom solution into a clean
beaker.
• Make sure the probe is ready for measurements (see
“Probe preparation” paragraph on page 10), i.e. the
membrane is filled with electrolyte and the probe is
connected to the meter.
• Switch the meter on by pressing the ON/OFF key.
• Remove the probe protective cap.