Do a Control Solution Test
Take a test strip out with clean and dry hands first.
Step 1 Insert Test Strip and Check
the Code Number
Be sure the code number on the dis -
play matches the code number on
the test strip vial. If the code numbers
do not match, please refer to “page
19~20” for the correct code selection.
Step 2 Press the M button
After blood drop symbol flashes, press
the M button and then “CTL” appears
on the display. With the “CTL” sign on
the display, the monitor will
NOT
store
your test result in memory.
Contact bars must be inserted all the way into the monitor or you may
get an inaccurate test result.
Every time you perform a control solution test, you must enter into the
“
CTL
” test mode so that the test result will not be stored in the monitor
memory. Failure to do so will confuse the blood glucose test result with
the control solution test result in memory.
WARNING!
63
23