Problem in Operation
1. Blood Glucose Measurement
What happened?
The monitor does not display a message after inserting a test strip.
What happened?
The test does not start after applying the sample.
What happened?
If you turned off the monitor but it makes a beeping sound.
POSSIBLE CAUSE
WH AT TO DO
Defective monito r.
Contact local customer service
Batteries exhausted
Replace the batteries.
Batteries incorrectly
installed or absent .
Check that the batteries ar e
correctly installed.
Test strip inserted upside
down or not completely
inserted.
Insert the test strip correctly
with the contact bars end first
and facing up.
POSSIBLE CAUSE
WH AT TO DO
The strip is still inside the slot .
R emove the strip. If you need to
test blood glucose, insert an
unused strip.
flashing
POSSIBLE CAUSE
WH AT TO DO
Defective monito r.
Contact local customer service
Insu fficient blood sample.
Repeat the test with a new test
strip and a larger sample.
Defective test strip.
Repeat the test with a new
test strip.
Sample was applied
.
Repeat the test with a new test
strip. Apply sample only when
appears on the display .
when flashing blood Symbol
was not displayed