Step 3 Obtain Control Solution
Shake the control solution vial well be-
fore squeezing. Discard the first drop,
wipe off the dispenser tip and squeeze
the second one into a fingertip.
Step 4 Apply Control Solution
Move your finger (with control solu
-
tion on it) close to the absorbent hole.
The drop is auto-drawn into the
test
strip. The monitor begins to count
down.
To avoid contamination,
please do not
directly apply
the control solution into
the test strip.
Step 5 Read and Compare the Re
-
sult
Compare the result with the range
printed on the test strip vial. The result
should fall within this range.
Out-of-range results
If test results fall outside the range printed on the test strip vial, check the
section of “Problem in Operation” in troubleshooting guide and repeat
the test. If you continue to get out-of-range results, it means that the
system may not be working properly. Do
NOT
test your blood. Contact
your local customer service for help.
109~165
Normal
BLOOD GLUCOSE MEASUREMENT