Appendix B: Error Codes
Advent Instruments Inc.
If the Ai5120Update.exe program has been
used at least once after the last calibration,
execute it again. It should be able to restore
the missing data. If not, please contact
technical support for assistance.
Very Fast
Flashing
Very Fast
Flashing
Very Fast
Flashing
Program Stack Fault
The AI-5120 software has encountered a
software stack fault. This error prevents the
AI-5120 from responding to any commands
and effectively freezes the system.
The power must be turned off and on in order
to restore normal operation. Please contact
technical support if this error occurs. It
represents a software problem and should be
corrected.
n/a
The following error code is a special case in which the "Ready" indicator is not flashing.
This error can only occur shortly after the AI-5120 is powered on.
READY
Indicator
RINGING
Indicator
OFF-HOOK
Indicator
Description / Corrective Measure
Error
Code
Off Off Off
Flash Memory Corruption
Failure to execute program:
The primary DSP application program could
not be started. This may indicate a corrupted
file within the flash memory, or a hardware
fault.
Use the Ai5120Update.exe program to restore
the flash memory contents. If this fails to
work, please contact technical support for
assistance.
n/a
26
AI-5120 User Guide