
Advanced Instruments, Inc
45
Adjust 4 mA with Zero O
2
Access the MAIN MENU by pressing the MENU key.
Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
Press the ENTER key to select the highlighted menu option.
The following displays appear:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
>>>
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT SPAN
OUTPUT ZERO
Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
Press the ENTER key to select the highlighted menu option.
The following display appears:
100.0
OUTPUT ZERO OFFSET
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
Enter the calculated adjustment value.
Note:
Once the initial adjustment is made and checked at the PLC
it may be necessary to fine tune the initial adjustment by repeating. Any additional percent error must be
added or subtracted from the initial adjustment value
000.0
OUTPUT ZERO OFFSET
PRESS UP OR DOWN
TO CHANGE VALUE
ENTER TO SAVE
MENU TO RETURN
Press the ENTER key to advance the underline cursor right or press the MENU key to advance the
underline cursor left to reach to the desired digit of the adjustment OUTPUT ZERO OFFSET value.
Press the ARROW keys to enter each the numerical value of each digit of the adjustment OUTPUT ZERO
OFFSET value.
Repeat until the complete OUTPUT ZERO OFFSET value has been entered.
Save the adjustment value by pressing the ENTER key or abort by pressing the MENU key.
The system returns to the SAMPLING mode.
Содержание GPR-2800 IS
Страница 6: ...Advanced Instruments Inc 5 2 Quality Control Certification ...
Страница 11: ...Advanced Instruments Inc 10 4 Features Specifications ...
Страница 48: ...Advanced Instruments Inc 47 ...
Страница 49: ...Advanced Instruments Inc 48 ...
Страница 50: ...Advanced Instruments Inc 49 ...