Advanced Instruments, Inc
27
CALIBRATION will appear and the analyzer will return to Sample mode without completing the Zero
calibration.
Both the Zero Calibrate and Span Calibrate functions result in the following displays:
PASSED CALIBRATION
OR
FAILED CALIBRATION
Default Zero
This feature will eliminate any previous zero calibration adjustment and display the actual signal output of
the sensor at a specified oxygen concentration. For example, assuming a zero gas is introduced, the display
above 0.000% will reflect an actual zero offset. This feature allows the user to ensure that the accumulative
zero offset never exceeds 50% of the lowest range limit. To perform Default Zero,
Access the MAIN MENU by pressing the MENU key.
Advance the reverse shade cursor using the ARROW keys to highlight CALIBRATION.
Press the ENTER key to select the highlighted menu option.
The following displays appear:
MAIN MENU
AUTO SAMPLE
MANUAL SAMPLE
CALIBRATION
CONFIG ALARMS
BYPASS ALARMS
>>>
CALIBRATION
SPAN CALIBRATE
ZERO CALIBRATE
DEFAULT SPAN
DEFAULT ZERO
OUTPUT SPAN
OUTPUT ZERO
Advance the reverse shade cursor using the ARROW keys to highlight DEFAULT ZERO.
Press the ENTER key to select the highlighted menu option.
The following display appears and after 3 seconds the system returns to the SAMPLING mode:
FACTORY
DEFAULTS
SET
0.03 %
AUTO SAMPLING
1% RANGE
76 F 100 KPA