43
Product Specifications
Appendix B
Electrical
Voltage:
Frequency:
Fuses (2):
Power:
Memory backup:
Sample Volume:
Sample Capacity:
Readout:
Units:
Range:
Linearity:
Resolution:
Repeatability:
Communications:
Storage
Temperature:
Operating Temp.
(ambient):
Drift:
100 to 130 VAC or 200 to 250 VAC
50 or 60 Hz
250V time delay (Type T): 1.25-Amp (100-130 VAC),
1-Amp (200-250 VAC)
100 Watts
Integral lithium cell; two years minimum
20 µL
Single sample
2-line digital display
mOsm/kg H
2
O
0 to 2000 mOsm/kg H
2
O
Within 1% of straight line between calibration points of 50
and 850 mOsm. Within 1.5% above 850 mOsm
1 mOsm/kg H
2
O
± 2 mOsm/kg H
2
O (1 S.D.) between 0 and 400 mOsm/kg
H
2
0; ± 0.5% (1 S.D.) between 400 and 2000 mOsm/kg H
2
O
DTE EIA-232/V.24 (RS-232), parallel printer, and optional
barcode scanner.
-40 to +160°F or -40 to +70°C
64 to 95°F or 18 to 35°C
Less than 1 digit per month plus 1 digit for every 9°F
or 5°C ambient temperature change
Содержание 3300
Страница 1: ...The Advanced Micro Osmometer Model 3300 User s Guide 3305 Rev9 012303 ...
Страница 6: ...vi Notes ...
Страница 8: ...viii Notes ...
Страница 12: ...xii Advanced Osmometer Model 3300 User s Guide Notes ...
Страница 18: ...xviii Advanced Osmometer Model 3300 User s Guide Notes ...
Страница 42: ...Advanced Osmometer Model 3300 User s Guide 24 Notes ...
Страница 54: ...36 Advanced Osmometer Model 3300 User s Guide Notes ...
Страница 68: ...Advanced Osmometer Model 3300 User s Guide 50 Notes ...
Страница 72: ...Advanced Osmometer Model 3300 User s Guide 54 Date Problem Symptom Action ...
Страница 75: ... Chapter Name ...
Страница 76: ... Advanced Osmometer Model 3300 User s Guide ...
Страница 77: ... Chapter Name ...