8. Tighten the calibration
setscrew with the calibra-
tion gauge.
9. Screw the calibration
gauge and key back into
the top of the sampler.
Your 20-µL sampler is now
calibrated and ready to use.
For verification that the wire
is calibrated correctly, use
the following procedure.
1. Place a new sample tip on
the sampler
2. Unscrew the calibration
gauge and key.
4. Insert the key end of the
calibration gauge into the
sample tip.
5. Visually inspect the posi-
tion of the end of the
sampler plunger tip and
the end of the calibration
key. There should be no
gap between the two.
6. If necessary, reset the
sampler plunger wire as
described above.
35
Figure 5:
Sample plunger replacement
and calibration
Troubleshooting & Service
Содержание 3300
Страница 1: ...The Advanced Micro Osmometer Model 3300 User s Guide 3305 Rev9 012303 ...
Страница 6: ...vi Notes ...
Страница 8: ...viii Notes ...
Страница 12: ...xii Advanced Osmometer Model 3300 User s Guide Notes ...
Страница 18: ...xviii Advanced Osmometer Model 3300 User s Guide Notes ...
Страница 42: ...Advanced Osmometer Model 3300 User s Guide 24 Notes ...
Страница 54: ...36 Advanced Osmometer Model 3300 User s Guide Notes ...
Страница 68: ...Advanced Osmometer Model 3300 User s Guide 50 Notes ...
Страница 72: ...Advanced Osmometer Model 3300 User s Guide 54 Date Problem Symptom Action ...
Страница 75: ... Chapter Name ...
Страница 76: ... Advanced Osmometer Model 3300 User s Guide ...
Страница 77: ... Chapter Name ...