Power frequency
(50/60Hz) mag
-
netic field
IEC 61000-4-8
3.0 A/m
3.0 A/m
Power frequency mag-
netic fields should be
at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC61000-4-11
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
<5% Ut (>95% dip in Ut)
for 0.5 cycle
40% Ut (60% dip in Ut)
for 5 cycles
70% Ut (30% dip in Ut)
for 25 cycles
<5% Ut (>95% dip in Ut)
for 5 sec.
Mains power quaility
should be that of a typical
commercial or hospital
environment. If the user
of the AMM213TD2 im
-
age intensifier requires
continued operation dur-
ing power mains interrup-
tions, it is recommended
that the AMM213TD2
image intensifier be
powered from an uninter-
ruptible power supply or
a battery.
Note: Ut
is the A.C. mains voltage prior to application of the test level.
The Model AMM213TD2 is intended for use in the electromagnetic environment specified below.
The customer or the user of AMM213TD2 should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance Level
Electromagnetic environment
- guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2.5 GHz
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2.5 GHz
Portable and mobile RF communications
equipment should be used no closer to
any part of the
AMM213TD2
, including
cables, than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation Distance
0.15 MHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
where
P
is the maximum oputput power
rating of the transmitter in watts (
W
)
according to the transmitter manufacturer
and
d
is the recommended separation
distance in meters (
m
).
Field strengths from fixed RF transmitters,
as deter-mined by an electromagnetic
site survey
(a)
, should be less than the
compliance level in each frequency range
(b)
.Interferency may occur in the vicinity
of equipment marked with the follwing
symbol:
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast, connot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the AMM213TD2 is used exceeds the applicable RF
compliance level above, the AMM213TD2 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the AMM213TD2.
(b) Over the frequency range 150kHz to 80MHz, field strengths should be less than [V1] V/m.
Manufacturer’s declaration - electromagnetic immunity
AMM213TD2
29