ST-LC OPERATION MANUAL(EN)    2011-04-27

 

7 / 35 

■ 

OPERATIONS 

 

■ 

FRONT PANEL 

 

 
 
 
 
 
 
 
 
 
 
 
 

 

EN

 

?

 

 

?

 

 

?

 

 

ENT 

  

  

  

MULTI-FUNCTION CONVERTER 

M.H.  D.H.

 

COM  RL 

   

Comm. status

 

Relay status 

Operation key 

Display screen 

 

„

 

Numeric Screen: 

 

 

0.28”(0.71cm) red

 

high-brightness LED

 

„

 

I/O Status Indicators: 

 

Relay Energized: 1~3 square read LED

 

 

RL1 

  display when Relay 1 has been energized

 

 

RL2 

  display when Relay 2 has been energized

 

 

RL3 

  display when Relay 3 has been energized

     

 

 

COM 

RS 485 Communiaction: 1 square orange LED 

will flash express the meter is receive or send data,

 

and flash quickly means the data transient quicker.

 

Stickers:

 

Each

 

meter with a sticker to describe what 

the functions together with engineer label 
enclosure. 

 

Function Stickers: 

 

 

 

Relay Energized:

 

 

H H 

 

HH Energized       

 

H i 

 Hi Energized 

 

L o 

 Lo Energized        

 

L L 

 LL Energized 

 

Hi. H 

  Hi Energized & Latch 

 

Lo. H 

  Lo Energized & Latch

 

D O 

 RS485 Energized         

 

„

 

Pass Code: 

Settable range:0000~9999;

 

User must key-in the exactly pass code for access 

to 

[

Engineer Level

]

. Otherwise, the meter will return to 

measuring page. If user forgets the pass code, please 
contact with your service window. 

„

 

Function Lock: 

There are 3 levels programmable.

 

 

None(

 none

):

 no lock at all. User can access to all 

level for checking and setting. 

 

User Level(

 user

): 

User Level lock. User can access 

to User Level for checking, but can not setting. 

 

Programming Level(

  eng

): 

Programming level lock. 

User can access to programming level for checking,but 
can not setting. 

 

ALL(

  all

): 

All lock. User can access to all level for 

checking but can not setting. 

 
 
 
 

■ 

Error Massage 

BEFORE POWER ON, PLEASE CHECK THE SPECIFICATION AND CONNECTION AGAIN. 

DISPLAY 

DESCRIPTION

:

 

REMARK 

 ovfl

 

 

Display is positive-overflow (Signal is over display 
range)

 

Please check the input signal 

-ovfl

 

 

Display is negative-overflow (Signal is under display 
range) 

Please check the input signal 

 ovfl

 

 

ADC is positive-overflow (Signal is higher than input 
120%) 

Please check the input signal 

-ovfl

 

 

ADC is negative-overflow (Signal is lower than input 
-120%) 

Please check the input signal 

  eep

 

 

 

fail

 

EEPROM occurs error 

Please send back to manufactory for 
repaired 

aiCng

 

 

 

   pv

 

Didn’t execute the calibrate of Input Signal 

Please process Calibrating Input Signal 

  aic

 

 

 

fail

 

Input signal calibrated error 

Please check Calibrating Input Signal 

aoCng

 

 

 

   pv

 

Didn’t execute the calibrate of output Signal   

Please process Calibrating Output Signal 

  aoc

 

 

 

fail

 

Output Signal calibrated error 

Please check Calibrating Output Signal 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Содержание ST-LC

Страница 1: ......

Страница 2: ...ample input 1 is 4 20mA and input 2 is Pt100 Optional This two inputs may also establish the related parameter individually establish mathematics function between two inputs Input Range aIlo aOhi The...

Страница 3: ...longer appear the input signal on the meter then the PV screen will display the data from RS485 command The data number will be same function as PV which it will compare with set point analogue output...

Страница 4: ...ilter band fixed in firmware and about 5 of stable reading 3 times Digital Filter set to be 3 continuously the meter will admit the samples and update the new reading Otherwise it will be as treat as...

Страница 5: ...alues Reverse slope output is decided by reversing point positions Output range selection aOtyp Voltage output specified Programming V0 10 0 10V V0 5 0 5V V1 5 1 5V Currnet output specified Programmin...

Страница 6: ...T MULTI FUNCTION CONVERTER M H D H COM RL E ENT MULTI FUNCTION CONVERTER M H D H COM RL Field Calibration In pass time the engineers have to take a lot of time to adjust and calibrate meters or conver...

Страница 7: ...deep Wiring Diagram Connector 10A 300Vac M3 0 1 3 3 5mm2 22 12AWG Max torque 13Kg cm 7 0mm max 7 0mm max Please set the torque of automotive screwdriver to match the limited of terminals OUTPUT 1 OUT...

Страница 8: ...ls programmable None none no lock at all User can access to all level for checking and setting User Level user User Level lock User can access to User Level for checking but can not setting Programmin...

Страница 9: ...nction index to access setting status 3 Setting Confirmed save to EEProm and go to next function index Shift key 1 In measuring page press for 1 second to access user level 2 In function index press f...

Страница 10: ...t for energized latch of Relay yes no 0 8 Mrst no Mrst Maximum Minimum storage reset yes no 16 88 MEASURING PAGE 1 1 0 3 ry sp 10 00 ry sp Relay 1 Set point setting Settable Range 19999 29999 0 6 miN1...

Страница 11: ...ized mode off lo hi lOhld hIhld do D 3 aOhs 10 00 aOhs Analogue High Output relative High Scale Settable Range 19999 29999 E 3 prity NstB2 prity Parity Nstb 1 Nstb 2 odd even A1 4 lOsC1 00 lOsC1 Low s...

Страница 12: ...lo trKhi B 9 ry rd 0 0 ry rd Relay 2 energized delay time Settable Range 0 00 0 9 m 59 9 s A1 10 dsply pV12 dsply Display Function pV12 miN12 maX12 rs485 A2 10 dsply pV12 dsply Display Function pV12 m...

Страница 13: ...lOsC1 or lOsC2 F2 or F7 and hIsC1 or hIsC2 F4 or F9 Please double check the lOsC1 or lOsC2 step A1 4 or A2 4 and hIsC1 or hIsC2 A1 5 or A2 5 are correct after selection the deflt or field Pass code de...

Страница 14: ...ish the calibration lower point and go to next page F 7 lOsC2 88888 lOsC2 the value to be set is relative to Field Calibration low point of input 2 Press to set the value of lower scale F 8 ca hi 8888...

Страница 15: ...table Range 19999 99999 Shift Up Down Enter 0 5 rYrst no no yes rYrst rY rSt Reset for energized hold of Relay If the ry_ md step B 3 7 set to be hIhld or lOhld and the present value PV reach to the c...

Страница 16: ...the pass code please go to step A1 15 or A2 15 to set Please remind the new pass code Pass Code inpT1 group inpT2 group relay group ao group rs485 group INPUT 1 GROUP INPUT 2 GROUP RELAY GROUP ANALOG...

Страница 17: ...o Adjustment for PV1 display For Zero Span of PV users can get the Fine Adjustment by front key It s an easy way to Just Key In the value that the user wants to show in the current input signal Especi...

Страница 18: ...ontrol analogue output relay energized and so on Programmable pV12 PV 12 Show PV 12 miN12 Min 12 Minimum Hold of PV 12 maX12 MAX 12 Maximum Hold of PV 12 rs485 RS485 Remote displayed from RS485 comman...

Страница 19: ...5 of stable reading 3 times Digital Filter set to be 3 continuously the meter will admit the samples and update the new reading Otherwise it will be as treat as a noise and skip the samples Settable...

Страница 20: ...Please refer to the explanations in INPUT SCALING of FUNCTION DEFINE page 1 35 Settable range 0 00 100 00 Shift Up Down Enter A2 3 pv dp 0 0 0000 pv dp Pv2 dP Decimal Point of PV1 Programmable 0 0 00...

Страница 21: ...pv sn Pv2 Sn Fine Span Adjustment for PV1 display It s same function as like as pv zo Settable range 19999 29999 Shift Up Down Enter A2 8 ZScl2 none none both ZScl2 Z S CL2 Clear Fine Zero Span Adjus...

Страница 22: ...vG Average The meter s sampling is 15cycle sec If the avg Average set to be 3 to express the display update with 5 times sec The meter will calculate the sampling 1 3 and update the display value At m...

Страница 23: ...not setting all All Level All lock User can access to all level for checking but can not setting Up Down Enter RELAY GROUP The group will not be displayed except the relay function is to be specified...

Страница 24: ...int hIhld lOhld High Low Level energized hold High Low Level energize and latch As the PV Higher or lower than set point the relay will be energized to latch except manual reset by from key in User Le...

Страница 25: ...AGE In following pages press for 1 second to back ANALOGUE GROUP INDEX PAGE D 1 aOtyp A4 20 A4 20 V0 10 aOtyp Ao tYP Analogue Output type Analogue output type had been fixed in mA or V as customer ord...

Страница 26: ...the front key up or down key to adjust and check the output of meter Settable range 38011 27524 Shift Up Down Enter D 5 aOspn 0 00000 32767 aOspn Ao SPn Fine Span Adjustment for Analog Output Users c...

Страница 27: ...lowing pages press for 1 second to back RS485 GROUP INDEX PAGE E 1 adres 1 001 255 adres AdrES Device number of the meter Settable range 1 255 Shift Up Down Enter E 2 baud 9600 9600 38400 baud bAUd Ba...

Страница 28: ...lo CA1 Lo Field Calibration Low of input 1 The low calibration is not need the exactly zero to calibrate because of the field calibration function could be calibrate any lower point Waiting for the v...

Страница 29: ...of machinery of input 2 to the lower signal output status or any lower status F 6 ca lo 0022h 052eh ca lo CA2 Lo Field Calibration Low of input 2 The low calibration is not need the exactly zero to ca...

Страница 30: ...120 input low limit A Please check the aIlo and aIhi in input group are correct or not B Please changes another meter that is matching in the field Display shows ovfl ovfl 3 To inspect whether did th...

Страница 31: ...t shown Check if the label of meter for detail specification A Please check the product number and output O P _____ description again for confirmation the relay output is specified or not B Please se...

Страница 32: ...bright but no reply A Please confirms the CHECH SUM program is correct B Please check the interval of each command has to over 3 5byte Reply wrong data from the meter 2 Check if the square orange LED...

Страница 33: ...12 0007h 19999 99999 The Maximum of PV 12 Low word 0 R RS485 0008h 19999 99999 PV 12 be written in by RS485 High word R W RS485 0009h 19999 99999 PV 12 be written in by RS485 Low word R W aOls 000Ah 1...

Страница 34: ...R W dsply 0024h 0 3 Display Mode 0 PV 12 1 Minimum Hold 2 Maximum Hold 3 RS485 0 R W lOcT1 0025h 19999 19999 Low Cut of PV1 0 R W lOcT2 0026h 19999 19999 Low Cut of PV2 0 R W avg 0027h 1 99 Average 5...

Страница 35: ...Digital Output 1 R W ry hy 0043h 0000 5000 Hysteresis of Relay2 0 R W ry rd 0044h 0000 5999 0 1second Energized Delay Time of Relay2 0 R W ry fd 0045h 0000 5999 0 1second De Energized Delay Time of Re...

Страница 36: ...he information contained in this manual ADtek reserves the right to make improvements to this document and or product at any time without notice No part of this publication may be reproduced stored in...

Отзывы: