ST-LC OPERATION MANUAL(EN) 2011-04-27
28 / 35
F
−
4
hIsC1
250)0
250)
0
2
50)0
hIsC1
(Hi.SC.1): High Scale relative
Field Calibration High of input 1
Settable range: -19999~29999
Shift
Up
Down
Enter
F
−
5
CseL1
deflt
deflt
field
CseL1
(C.SEL.1): Calibration parameter
selection of input 1
As the user finished the procedures of field
calibration, the field calibration datum has
been saved in EEProm and it can't change
the default(factory) calibration datum . Even
the field calibration has been done, the user
can still select either default calibration or
field calibration.
Programmable:
deflt
(default):
factory calibration
points and factors
field
(field):
field calibration points
and factors
Up
Down
Enter
►
If the user select field calibration
of the
[
lOsC1
]
(step A1-4) and
[
hIsC1
]
(step A1-5) will be
replaced by
the
[
caLls
]
and
[
caLhs
]
which it
can not to be
change by anyone. If user has to
change the scaling, it's the only
way to access field calibration
level to set in
[
lOsC1
]
(step F-2)
and
[
hIsC1
]
(step F-4).
►
Please double check the
[
lOsC1
]
(step A1-4) and
[
hIsC1
]
(A1-5)
whether
are correct after selection
the
deflt
or
field.
Adjust the structure of machinery of input 2 to the lower signal output
status(or any lower status).
F
−
6
ca@lo
0022h
052e
h
ca@lo
(CA2.Lo): Field Calibration Low of
input 2
The low calibration is not need the exactly
"zero" to calibrate because of the "field
calibration" function could be calibrate any
lower point.
►
Waiting for the value till stable,
pressed
Key to read signal low
of sensing device.
►
Waiting for above reading stable
(around 3~5seconds), press
Key again to complete the
calibration lower point, and go to
[
lOsC2
].
F
−
7
lOsC2
)0
000)
0
0
00)0
lOsC2
(Lo.SC.2): Low Scale relative Field
Calibration Low of input 2
Settable range: -19999~29999
Shift
Up
Down
Enter
Adjust the structure of machinery of input 2 to the higher signal output
status(or any higher status).
F
−
8
ca@hi
6478h
699b
h
ca@hi
(CA2.Hi): Field Calibration High of
input 2
The high calibration is not need the exactly
"span" to calibrate because of the "field
calibration" function could be calibrating any
higher point.
►
Waiting for the value till stable,
pressed
Key to read signal high
of sensing device.
►
Waiting for above reading stable
(around 3~5seconds), press
Key again to complete the
calibration higherer point, and go to
[
hIsC2
]
。
F
−
9
hIsC2
250)0
250)
0
2
50)0
hIsC2
(Hi.SC.2): High Scale relative
Field Calibration High of input 2
Settable range: -19999~29999
Shift
Up
Down
Enter
Next Page