- 22 -
All analog outputs of the instrument are ACTIVE and SINGLE ENDED type, therefore
they can be connected only to PASSIVE receiver devices. The minimum load allowed for
voltage outputs is 10 kohm, the maximum load allowed for current outputs is 300 ohm.
SERIAL COMMUNICATION SETTING
-
/
: communication port.
-
: it disables any type of communication (default).
-
: MODBUS-RTU protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
-
: ASCII bidirectional protocol; possible addresses: from 1 to 99 (see Communication
protocols manual).
-
-
-
: continuous weight transmission protocol (see Communication protocols manual),
at the frequency set in
item (from 10 to 300).
-
-
,
-
: continuous weight transmission protocol to RIP5/20/60, RIP50SHA, RIPLED series
remote displays; the remote display shows the net weight or gross weight according to its
settings (set:
-
displays; the remote display shows the net weight or gross weight according to its settings
(set:
-
: continuous weight transmission protocol to RIP675, RIP6125C series remote
displays (set:
).
When the remote display is set to gross weight:
-
if the instrument displays the gross weight, the remote display shows the gross weight.
-
if the instrument shows the net weight, the remote display shows the net weight
alternated with the message
.
-
: printer.
ACTIVE
ANALOG
OUTPUT
PASSIVE
RECEIVER
Voltage or current analog signal
-
COMMON
+
ADN
Pesage
Transmetteur de pesage T16F : Manuel d’utilisation et d’installation