![background image](http://html1.mh-extra.com/html/admesy/mse-series/mse-series_operating-manual_2852906019.webp)
19
The test patterns consist of two sets, each containing 24 colours:
Unsaturated set (figure 6)
Saturated set (figure 7)
Fig 6 Unsaturated colours.
Fig 7 Saturated colours.
The results of auto-ranging show a stability level of approxi-0.001
on color. A higher adjmin level will logically improve this, though it
increases the total measurement time. Note that the wait time after setting
the pattern is 200ms for both tests. This wait time is display dependent.
The following table shows the timing results of measuring all 24 colours
(so, the total time for 24 patterns is shown). This time does not include the
waiting time (200ms) for the pictures to become stable, so this is purely
the real measurement time and communication time of MSE.
Initial integration time
Total measurement time
200
1.8
1.000
1.8
16.666 (1 frame)
1.7
30.000
1.8
50.000
2.8
100.000
5.5
200.000
10.5
Feedback
3.2
Table 13 Unsaturated colours.
The
bold
values indicate the initial integration time caused clipping. To
adjust down from clipping can take a long time since the MSE cannot tell
what the correct level is that will not clip. This situation should be avoided
as much as possible.