28
Operation Theory
3.4.3
Post-Trigger or Delay-Trigger Acquisition with Re-
Trigger
Use post-trigger or delay trigger acquisition with re-trigger function
to collect data after several trigger events, as illustrated in Figure
3-9. You can program the number of triggers then the digitizer will
acquire a specific sample data each time a trigger is accepted.
Thus the time between last sampled data and next trigger event
can be only one clock period of Timebase. After the initial setup,
the process does not require software intervention.
Figure 3-9: Re-Trigger Mode Operation