52
•
Calibration
7.2 VR Assignment
There are eight variable resistors (VR) on the PCI-9812/10 board to
allow you to make accurate adjustments on A/D channels. The
function of each VR is specified in Table 7.1.
VR1
A/D channel 0 offset adjustment
VR2
A/D channel 1 offset adjustment
VR3
A/D channel 2 offset adjustment
VR4
A/D channel 3 offset adjustment
VR5
A/D channel 0 full scale adjustment
VR6
A/D channel 1 full scale adjustment
VR7
A/D channel 2 full scale adjustment
VR8
A/D channel 3 full scale adjustment
Table 7.1 Function of VRs
7.3 A/D Calibration
7.3.1
AD Calibration for Channel 0
1. Apply a +1V input signal to A/D channel 0, and trim the
VR5
to
obtain the range of the average reading of channel 0 is within
2046.6
±
0.1 (9812) or 510.9
±
0.1 (9810).
2. Apply a +0V input signal to A/D channel 0, and trim the
VR1
to
obtain the range of the average reading of channel 0 is within
±
0.2.
3. Repeat step 1 and step 2, adjust
VR5
and
VR1
.
7.3.2
AD Calibration for Channel 1,2,3
You could calibrate other channels with the same procedure in
section 7.3.1 while trimming the corresponding VRs. Refer the
following table to adjust the correct VR.
Channel
Number
Full Scale
Adjustment VR
Offset
Adjustment VR
0
VR5
VR1
1
VR6
VR2
2
VR7
VR3
3
VR8
VR4
Содержание NuDAQ PCI-9812/10
Страница 1: ...NuDAQ PCI 9812 10 20MHz Simultaneous 4 CH Analog Input Card Users Guide ...
Страница 4: ......
Страница 57: ...C C Library 49 ...
Страница 61: ...Calibration 53 ...
Страница 67: ...Software Utility 59 For example if you select 3 the following figure will be displayed on the screen ...
Страница 69: ...Software Utility 61 ...