ASM 192 T2D+
Technical characteristics
Measurement range
(Helium)
Crossover pressure
(at inlet)
mbar.l/s
Pa.m
3
/s
mbar
Pa
Gross leak test mode
1.10
-5
to 1
1.10
-6
to 1.10
-1
30
3000
Normal test mode
1.10
-9
to 1.10
-1
1.10
-10
to 1.10
-2
3
3000
High sensitivity test mode
1.10
-11
to 3.10
-4
1.10
-12
to 3.10
-5
2.10
-2
2
Sniffing test mode
1.10
-7
to 1.10
-1
1.10
-8
to 1.10
-2
sniffer probe at atm. pressure
Response time (Inlet port blanked off)
< 0.1 s
Helium pumping speed:
At inlet port
20 l/s
Roughing (primary) pump characteristics:
Hybrid (primary) pump pumping speed (in air) (1 ACP 28 version)
27 m
3
/h (16 cfm) + 10 l/s
Hybrid (primary) pump pumping speed (in air) (2 ACP 28 version)
54 m
3
/h (32 cfm) + 10 l/s
Analyzer cell (Spectro):
Analyzer cell design
self protected 180° magnetic deviation mass spectrometer
Analyzer cell filament
2 tungsten filaments
Analyzer cell sensitivity
3.10
-4
A/mbar
Emission current range
0,2 to 2 mA
Air pumping speed
110 l/s
Helium pumping speed
30 l/s
Displays and setpoints adjustments:
Inlet port pressure display range
10
3
to 10
-3
mbar / 10
5
to 10
-1
Pa
Audio alarm:
100 dB frequency modulated and adjustable audio signal
Hard vacuum Audio signal set point
1.10
-19
to 9.9.10
19
Sniffing Audio set point
1.10
-19
to 9.9.10
19
Alcatel Vacuum Technology France - User's Manual ASM 182/192 TD+ - ASM 192 T2D+