A 200
ASM 182 TD+ / 192 TD+
detector operating principle
Pumping capacities
25 m
3
/h (15 cfm) roughing (dry primary pump ACP 28)
+ 10 l/s (molecular drag pump MDP 5011).
Helium pumping speed at inlet port: 4.4 l/s.
Test capacities
Short test cycle.
Quick response time.
Auto-calibration with integrated calibrated leak.
ASM 182 TD+ with gas line option
C 420
1
Preamplifier
2
Analyzer cell
3
Detector inlet port
4
Internal calibrated leak
5
Inlet pressure gauge (PI3C)
6
Outlet pressure gauge (PI1)
7
Vent connector
8
Long distance sniffer connector
9
Nitrogen purge connector
15
Roughing primary pump
(ACP 28)
16
Roughing molecular pump
(MDP 5011)
18
Detection turbomolecular pump
(TMP 5154)
V
A1
Inlet vent valve
V
R1
Roughing valve
V
B
By-pass valve
V
T1
Exhaust valve
V
T4
Detection valve
V
S
Sniffing valve
V
C2
Calibration valve - Detection
V
C1
Calibration valve - Roughing
V
C3
Calibration valve - Inlet vent
Vacuum circuit
12
1
2
18
V
S
8
6
4
V
C2
V
C3
V
C1
V
T1
V
B
9
15
16
5
V
R1
V
A1
7
3
V
T4
VAT valve
Bacosol valve
Minisol valves
Reference correspondence between valve/vacuum block marks
E 530
Alcatel Vacuum Technology France - User's Manual ASM 182/192 TD+ - ASM 192 T2D+
Содержание ASM 182 TD PLUS
Страница 1: ...ASM 182 TD ASM 192 TD ASM 192 T2D HELIUM LEAK DETECTOR User s Manual ...
Страница 4: ......