5.7 SENSe Subsystem
5.7.1 FREQuency Subsection
•
Set start frequency
Command format:
SENSe:FREQuency:STARt <freq>
SENSe:FREQuency:STARt?
Function description:
Set start frequency
Query start frequency
Parameter specification
freq, integral data, range 0Hz to 1.6GHz
Return format
Query and return start frequency as an integer, unit is Hz.
•
Set center frequency
Command format:
SENSe:FREQuency: CENTer <freq>
SENSe:FREQuency: CENTer?
Function description:
Set center frequency
Query center frequency
Parameter specification
freq, integral data, range 0Hz to 1.6GHz
Return format
Query and return center frequency as an integer, unit is Hz.
•
Set stop frequency
Command format:
SENSe:FREQuency: STOP <freq>
SENSe:FREQuency: STOP?
Function description:
Set stop frequency
Query stop frequency
Parameter specification
freq, integral data, range 0Hz to 1.6GHz
Return format
Query and return stop frequency as an integer, unit is Hz.
•
Set span frequency
Command format:
SENSe:FREQuency: SPAN <freq>
SENSe:FREQuency: SPAN?
Function description:
Set span frequency
Query span frequency
Содержание DSA2016TG
Страница 1: ...USER S MANUAL Handheld Spectrum Analyzer DSA2016TG V1 0 0 ...
Страница 7: ...Chapter1 Quick Start u Instrument Introduction u General Check u Preparation ...
Страница 15: ...2 3 2 Left Panel Left Panel 5 Headphone Connects to a headphone 6 SD card slot 5 6 ...
Страница 19: ...Note in the storage function the direction keys can also be used to select the current path or file ...
Страница 66: ...input 220VAC 15 AC frequency range 40Hz to 60Hz Power consumption Maximum 32W ...
Страница 67: ...Chapter7 Appendix u Appendix A u Appendix B ...